8/9/2019 Presentation ORC
1/21
NATURAL GAS MEASURMENT (CONT’D)
The method of measuring liquid or gas by
volume is referred to as volumetricmeasurement.
Producers account for gas in units of mcf(1000 cu.ft)
The total mass of a substance in a cu ft of gasdepends partly on its absolute pressure (psia)and partly on temperature (oR).
The American Petroleum Institute (API) and
American Gas Association (AGA), since 1967,have used 14.73 psia and 60of (520oR) as theirstandard (base) conditions.
-
8/9/2019 Presentation ORC
2/21
ORIFICE METER
This is a type of differential pressure method of meteringgas
The orifice meters compound of two elements
• Primary Element
• Secondary Element
-
8/9/2019 Presentation ORC
3/21
PRIMARY ELEMENT
The primary element is the differential pressure-
producing device. It composed of the followingparts:
The meter tube – a length of special pipe through which the gas flows.
The orifice plate holding and positioning device-an orifice flange or an orifice filting installed as anintegral part of the meter tube to hold the orificeplate in a position perpendicular and concentric to
the flow of gas. The orifice plate- a flat circular plate with acentrally bored, sharp-edged orifice machined toan exact.
-
8/9/2019 Presentation ORC
4/21
Predetermined dimension that forms acalibrated restriction to the flow of gas throughthe meter tube and is the source of thedifferential.
Pressure taps – Precisely located holesthrough the pipe walls or orifice plate may bemeasured.
Straightening vanes – a device that may beinserted in the upstream section of the metertube swirling in the gas stream.
-
8/9/2019 Presentation ORC
5/21
SECONDARY ELEMENT
The secondary element is called the differential
gauge and is connected to the upstream anddownstream pressure taps of the primaryelement.
It measures the differential pressure and static
pressure which are recorded on a circular chart.
-
8/9/2019 Presentation ORC
6/21
BASIC ORIFICE METER EQUATION
From the general energy conservation equation
for any points in the meter tube
∫ ∫ ∫ −−=++2
1
2
1
2
1lwwdz
gc
g
gc
udvVdp
Where,
v = sepcific volume = cu ft/ibmp = pressure, ibf/sq ftu = average linear – flow velocity ft/secgc = conversion factor: 32.17 (ibm/ibf) (ft/sec
2)
g = acceleration due to gravity ft/sec2z = vertical distance above datum, ft
w = work done by flowing fluid, ft ibf/ibmLw = work energy lost due to frictional effect, ft-
ibf/ibm.
-
8/9/2019 Presentation ORC
7/21
BASIC ORIFICE METER EQUATION (CONT’D) From the assumption, dx = o, w = o
O gcduU Vdp =+∫ ∫
2
1
2
1
From the above equation, it can be shown that,
−
∆
=
z T
hp DC
P
T q
g b
b
γ β 41
59.2182
2
Therefore,
f w g phC q 1=
-
8/9/2019 Presentation ORC
8/21
GENERALIZED ORIFICE FLOW EQUITATION
The generalized orifice flow equation is given as:
f wh P hC q
1=
f wT h
Where,qg is the gas flow rate, scf/hr
h w is the differential pressure (in inches of water @ 60o
f)pf is the static pressure (in psia)
CI is the orifice flow constant
Is the pressure extension.
This constant is a function of several other orificeconstants given by:
CI = (F b) (Fr) (Y) (Fpb) (Ftb) (Fg) (Fpv) (Fm) (Ft) (Fa)
-
8/9/2019 Presentation ORC
9/21
GENERALIZED ORIFICE FLOW EQUATION (CONT’D)
Where
F b = Basic orifice factor
Fr = Reynolds number factor (viscosity)
Y = Expansion factor
Fpb = Pressure base factor (14.73) contracted pressure base
Ftb = Temperature bases factor (T b/520)
-
8/9/2019 Presentation ORC
10/21
Ftf= Flowing temperature factor
Fg= Specitic gravity facator,
Fpv = Super compressibility factor
Fm = Manometer factor for me-cury meter.
Ft= Guage location factor
Fa = Orifice thermal expansion factor
= 1 + [0.000,018,5 (of – 68)] 304 and 316 stainless steel
=1 + [0.000, 015, 9 (of – 68)] monel
etemperatur flowactual +400/520
g γ 1
z
Zb
-
8/9/2019 Presentation ORC
11/21
QUESTION ON METERING Given an orifice meter equipped with flange taps withstatic pressure from the down stream taps and the
following:D1= line size = 8.071 inches of actual 1D
D2 = orifice size = 1.000 inch
Flowing temperature = 65of
Ambient temperature = 70
o
fP b = contract pressure base = 14.65 psia
Temperature base = 50of = 510.oRγg = Specific gravity = 0.570
H w
= total heating value = 999.1Btu cu.ft
Xn = mole fraction nitrogen content = 0.011
Xc = mole fraction of carbon (iv) oxide content = 0.000
H w average differential head = 50 inches water
Pf= average downstream gauge pressure = 370 psia Calculate the orifice flow constant and the uantit of
-
8/9/2019 Presentation ORC
12/21
0008.1
1239.01301.0/int
),(25.
0005.149000.01
64.138
0680.011
0680.0
1071.81,23.
38.200
1071.81,22.
64.13865.1437050
1239.0071.8
000.1
2
1
2
=
==
=+=
+=+=
=
−−
=
−−
=+=
===
γ
β
γ
β
and ph for erpolating by
pressure staticdownstream for Table !rom
ph
b !
b
line Dinanin plateina for Table !rom
!
line Dmanin plateina for table !rom
" ph
D
D
f w
f w
r
b
f w
-
8/9/2019 Presentation ORC
13/21
)()()()()()()()()()()(
tan
9993.0
)36.(int
,570.070,370
054.1),(35.
1,35.
3245.1
570.0,34.
9952.0
65,33.
9808.0
50,32.
0055.1
65.14,31.
1
at m pv g tf tb pbr b
m
g
o
f
pr
pv
g
tb
o
tb
o
pb
b
! ! ! ! ! ! ! ! # ! ! C
ist cons floworificetheThen
!
Tableerpolationby
and f etemperatur ambient psig P !or
! a Table !rom
! Table !rom
!
gravity specific for Table !rom
!
f etemperatur flowing for Table !rom
!
f baseetemperatur for Table !rom
f
P for Table !rom
=
=
===
=
==
==
==
=
==
γ
-
8/9/2019 Presentation ORC
14/21
hr $scf
hr ft
"
phC qh
hour one for flowof ratetheThrn "
C
f w
/05.37
/052,37
4.3845025.267
,25.267)9993.0()0254.1()3245.1()9952.0(
)9808.0)0055.1()0008.1()0005.1()38.200(
3
===
=
=
=
-
8/9/2019 Presentation ORC
15/21
CHART
Two principal types of chart are:
Uniform scale direct – reading chart for differential
pressure – line are spaced on equal distance
Square – root chart – reads the square root of the
percentage of the full scale range of the gauge.
DIRECT – READING CHART
Circular charts for recordity the differential andstatic pressure guages are usually 12 in. in
diameter
Scale ranges for fuel gas metering are given below.
-
8/9/2019 Presentation ORC
16/21
Common differential pressure ranges, in.
of water
Common static pressure ranges,
psig
0 to 10
0 to 20
0 to 50
0 to 100
0 to 200
0 to 100
0 to 250
0 to 500
0 to 1000
0 to 2500
-
8/9/2019 Presentation ORC
17/21
SQUARE – ROOT CHART For this type of chart, a chart factor (CF) may bedefined as:
100
$%C! =
Actual pressure
rangemeter "C%
C! "C%
2
2
10
)(
=
=
Where, MR = Meter Range CR = Chart Reading
-
8/9/2019 Presentation ORC
18/21
EXAMPLE ON CHART Given that for a 50in by 100 Ib gauge, the differential pressuregauge Rh = 50 in and static pressure; Rp = 100 psi. assume square
root chart readings; differential = 7.2 in and static = 9.4 psi.Calculate the actual differential and static pressure.
SOLUTION
psia "
% "reading chart
pf
pressure&tatic
water in "
% "reading chart
h
pressureal Differenti
p
hw
36.8810010
4.9
10
.92.255019
2.7
10
,
2
2
2
2
=
=
=
=
=
=
-
8/9/2019 Presentation ORC
19/21
VENTURI METER
The venturi tube operates on exactly the same
principle as the orifice meter. Discharge coefficients of venturi are larger than
those for orifices and vary from about 0.94 to
0.99. It gives a definite improvement in power losses
over an orifice and is often indicated for
measuring very large flow rates where power
losses can become economically significant.
The initial higher cost of a venturi over an orifice
may thus be offset by reduced operating costs.
-
8/9/2019 Presentation ORC
20/21
INSTRUMENT CALIBRATION Calibration entails the adjustment of a measurementdevice so that the value from the measurement device
agrees with the value from a standard. The international science organization (ISO) hasdeveloped a number of standards specially directed tocalibration of measurement devices. Within most companies, the responsibility for
calibration measurement device is delegated to a sspecific department. The frequency of calibration is normallypredetermined but earlier action may be dictated if
the values from the measurement device becomesuspected. Calibration of some measurement devices involvescomparing the measured value with the value fromthe working standard.
-
8/9/2019 Presentation ORC
21/21