mustafa_thesis presentation
TRANSCRIPT
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg1
Master Thesis :The clinical performance of the DAVID-system for the
in vivo verification of VMAT irradiation
Presented by Mustafa Saibu Danpullo
1st supervisor Prof. Dr.B.Poppe
2nd supervisor Dr. HK. Looe,
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg2
Layout
I IntroductionII Theory
VMAT and IMRT MLC Design and Agility MWPC and DAVID system
III Materials & Methods Equipment, alignment, patient data, stability of DAVID chamber Beam property, Error detection Deconvolution DAVID QA software
IV Results / DiscussionV Conclusion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg3
I Introduction
IMRT (Intensity-modulated radiation therapy)VMAT (Volumetric Modulated Arc Therapy):
Why In vivo verification ?
ICRU report 24 (1976) recommended that certain types of tumors requires improve accuracy from 5% to 3.5%.
To detect Equipment-related errors and deviations from the initial plan Complexity of planning and delivery techniques increases risk for
treatment-related error incidents. In 1992 to 2007, more than 4,000 near misses without adverse
outcome to patient’s case were reported, more than 50% were related to the planning or treatment delivery stage.
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg4
In vivo dosimetry methods: in vivo intracavitary dosimetry with TLD Diodes
DAVID (Device for Advanced Verification of IMRT deliveries In-vivo verification during treatment
Online measurement of differences in dose to reference Error detection of the Multi Leaf Collimator (MLC)
I Introduction
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg5
II Theory
Mostly Siemens, Elekta and Varian have introduced new LINAC control systems that will be able to change the MLC leaf positions
IMRT uses many small fields to generated by beam-shaping devices (MLC) to deliver a single dose of radiation
IMRT: Intensity-modulated Radiation Therapie
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg6
II TheoryVMAT : Volumetric Modulated Arc Therapy
VMAT is a rotational IMRT that can be delivered using conventional LINAC with MLC
Elekta and Varian have introduced new LINAC control systems that will be able to change the MLC leaf positions and dose rate
while the gantry is rotating.Precise Beam infinity and Rapid Arc
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg7
A schematic drawing of the Siemens type A, Elekta type B and Varian type C MLC [18]
Stepped leafs for different manufacturers [34]
II Theory
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg8
II Theory: Agility MLC design
• 160 tungsten leafs, • rounded arc edge, • 5 mm width, • High speed(2x normal MLC) of up to
3cm/sec,• large field MLC enable clinicians to
shape radiation, • extremely low transmission of about
<0.5% • 45 cm isocenter clearance from
accessory holder.
MLC Motor
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg9
Inventor:Prof. Georges Charpak, France,1968
Nobel Prize in Physics (1992)
II Theory: Multi wire proportional chamber (MWPC)
The DAVID chamber is a multi-wire ionization chamber designed by PTW Freiburg based on Charpark's multi wire proportional chamber.
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg10
Compton scattering Electric field causes electrons move to the anode(wire) and ionizied atoms/molecules to the cathode(plate)
Each detection wire accumulates charge which loads a C.
After the voltage at the capacitor is read out, it is set to zero and charged again
The voltage achieved is read out by the associated amplifier at a rate of 1 Hz.
Performed by multi-channel electrometer (MULTIDOS) + additional Software
.
II Theory: DAVID system functioning principle
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg11
Signal interpretation:
Ri: reading of a single channel (ion charge collected)C: cross section of the lengthy collection volume along the wireIi: ionization density (x1 start of wire, x2 end of wire)
li1-li2: aperture of the associated leave pair
II Theory
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg12
Front Plate
Back Plate
Air Volume
II Theory: DAVID system signal recording
3 groups of secondary electrons contributing to the signal:a)“primary signal”
b)scattered signal
c)leakage radiation (background signal)
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg13
VMAT Planning:Treatment Planning System: ONCENTRA Masterplan Version 4.3ELEKTA Synergy accelerator with an Agility 80 leaf-pair MLC• Desktop Pro TM 7.011 is Elekta's third generation fully integrated
digital control system. MOSAIQ, DAVID software version 2.0 DAVID T34065
III Materials and Methods
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg14
Patient data configuration chart
III Materials & Methods
Patient data
Reference
1st session
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg15
DAVID Analysis• PTW: DAVID 2.0 software
III Materials & Methods
Warning level: 3%
Alarm level: 5%
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg16
III Materials & Methods:
• VMAT: 4 (1 H&N, 3 Prostates)• 180° to -180° Clockwise and anti clockwise
Stability of the DAVID system
• IMRT : 1 (Prostate)• 0°• 90°• 270°
(Prostate and Head and Neck) 14 days
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg17
III Materials & Methods: The beam property of the DAVID chamber
• Percentage depth dose (PDD)• Roos chamber 34001 • MP3 water phantom
• Transmission factor for 6 and 15 MV• Semifex T31010 (Diff Field sizes)
Setup conditions
• With and without DAVID• SSD 100 and 80 cm• Photon energy 6 and 15
MV• Different field sizes
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg18
1. Successive opening of 1 leaf on 1 side
III Materials & Methods:VMAT Plan Editing for error detection
• MATLAB script to change the MLC-positions
3. Field shift of a leaf gap (size of leave gap remains)
2. Successive shift of a leaf gap (size of leave gap remains)
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg19
1. Successive opening of 1 leaf on 1 side
VMAT Plan Editing for error detection • MATLAB script to change the MLC-positions
3. Field shift of a leaf gap (size of leave gap remains)
2. Successive shift of a leaf gap (size of leave gap remains)
III Materials & Methods:
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg20
DeconvolutionIII Materials & Methods:
S(x) measured signal as convolution of P(x) True „dose“ profile with LRF fɛ(x).
S(x) = P(x) * f(x) „van Cittert“ iterative deconvolution algorithm
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg21
opened MLC at every 10th interval from 1st to 80th pairs.
Nine MLC slit through the entire DAVID chamber
IV Results and Discussion: Alignment
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg22
IV Results and Discussion: Stability of DAVID System
• IMRT: prostate• Deviation of ±2% (+2%)
• VMAT: prostate• Deviation of ±1% (-1%)
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg23
IV Results/Discussion: Stability of DAVID System
• IMRT: prostate• Deviation of ±2% (+2%)
• VMAT: H&N• Deviation of <0.5%
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg24
IV Results/Discussion: Transmission factor
The average KDAVID • 0.939 ±0.003 for 6 MV
and• 0.953 ± 0.004 for 15 MV
Reduction of dose at isocenter due to 8mm of PMMA
By measuring the attenuation factor the output value can be corrected.
Attenuation of the beam by the DAVID chamber
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg25
100 cm SSD,Increased about 0.32% with DAVID No change of Dmax 1.4 cm with and without DAVID
80 cm SSD,An increased about 4.26% with DAVID slight change of Dmax 1.3 cm with DAVID 1.4 cm without DAVID
IV Results/Discussion: Changes in PDD
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg26
100 cm SSD,An increased about 0.67% with DAVID Dmax 2.6 cm No change
80 cm SSD,An increased about 18% with DAVID slight change of Dmax 1.8 cm with DAVID 2.4 cm without DAVID
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg27
Surface dose Increase with increase in field sizeIncrease with increase in energyIncrease with decrease in SSD
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg28
Pronounce withdecrease in SSD and, increase in photon energy andincrease in field size.
Increase in surface dose is due to scattered secondary electrons from the DAVID
chamber reaching the water phantom surface
(electron contamination).6 cm
45 cm
100 cm
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg29
Deconvolution test by iteration method
IV Results/Discussion: Deconvolution
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg30
Deconvolution• does not depend on the length of the slit. • 40 cm slit results showed a small decrease in the tail signals.
10 x 10 cm
10 cm slit
40 cm slit
20 cm slit
30 cm slit
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg31
IV Results/Discussion:
Deconvoluted slit signal at 40th and 65th wire
Deconvolution test
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg32
10x10cm fields before and after deconvolution
Single arc prostate plan before and after deconvolution (3 mm )
IV Results/Discussion: Deconvolution with DAVID software.
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg33
before deconvolution 6.1 mm Gradients of the linear fit
before deconvolution : 0.47 and
after deconvolution 2.9 mm
Gradients of the linear fit after deconvolution : 0.94.
IV Results/Discussion: Enhance sensitivity after deconvolution
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg34
before deconvolution after deconvolution H&N 6mm error..
IV Results/Discussion
False alarm/warning effect before deconvolution.
The effect is eliminated after deconvolution
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg35
Undeconvoluted
deconvoluted
Measuring the deconvolution matrix with the DAVID software as manual. LSF of single middle slit is measured and
used to generate the 80x80 LRF matrix with MAT LAB and installed in the DAVID software for deconvolution.
IV Results/Discussion: Limitations of DAVID-160 system
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg36
2mm MLC error, single bank shift 2mm MLC error, single leaf
Analyzing both the maximum deviation and total deviation in two different plots at the same time
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg37
Max Dose 75.99GyMax Dose 57.27Gc
IV Results/Discussion: Artificial MLC bang shift Error 2 cm
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg38
0 8 10 12 14 16 18 200
1
2
3
4
5
6
7
8dose in-crease / Gy
maximal deviation / %
shift / mm
Undetectable error! Design of the chamber
DAVID: Gap-Shift (prostate with OAR: rectum back wall) IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg39
DAVID Quality Assurance software (DQA)
MAT LAB 2011b and 2012b analyzes the daily session for all patient's data with 2 clicks online.
• NDD- Non deconvoluted data• DD-Deconvoluted data�• ED- Electrical data
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg40
Analysis only specific data on specific date, session and only print out the deconvoluted data (DD)
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg41
Patient text file from DQA software
Sample patient 2014-01-12
Display only data's with MLC error indicating the data type (DD),Beam
number, Session, segment and particular MLC with error.
IV Results/Discussion:
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg42
The DQA software displays the warning and alarm errors
Warning and error dialogs at future date entry and invalid date entry respectively
year-month-day `yyyy-mm-dd'
IV Results/Discussion
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg43
VMAT and IMRT plans sessions
IV Results/Discussion:
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg44
DAVID chamber:• Linear dependency on leaf opening• Sensitivity dependent on leaf gap opening• How much radiation pass through the opening• Deconvolution double the sensitivity
DAVID is design for specifics LINACS Independent from the LINAC In-vivo verification of MLC malfunction during VMAT • Undetectability field shifts due to chamber design
(Suggestion: perpendicular wires or gradient)
V Conclusions
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg45
V Conclusion
Single MLC bank shift error Maximum and total deviation to be analysedDeconvolution matrix To be generated for each linac To be generated by single single slit• In comparison to other techniques measurement of undisturbed
signals-> no dependence on patient position(EPID)-> measurement of the complete delivered dose(TLD, diode or MOSFET detectors)• Suggestions for future development: Standard design for software
• Deconvolution program to be integrated,• DQA to be integrated • Design two chambers perpendicular to each other Co-operate directly with LINAC vendors for specifics designs
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg46
Sources
1. [1] Ezzel GA, Galvin JM, Low D, Palta JR, Rosen I , Sharpe MB, Xia P, Xiao Y, Xing L and Yu CX. Guidance on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT subcommittee of the AAPM radiation therapy committee Med Phys 2003; 30:2089-115.
2. [2] ESTRO Booklet No. 9, 2008. Guidelines for the verification of IMRT, edited by Ben Mijnheer and Dietmar Georg. ISBN 90-804532-9.
3. [3] Schneider F, Polednik M, Wol D, Steil V, Delana A, Wenz F, Menegotti L. Optimization of the gafchromic EBT protocol for IMRT QA. Z Med Phys 2009; 19(1):29-37.
4. [4] Poppe B, Blechschmidt A, Djouguela A, Kollho R, Rubach A and Harder D. Two-dimensional ionisation-chamber arrays for IMRT plan verication. Med Phys 2006; 33:1005-15
5. [5] Poppe B, Thieke C, Beyer D, Kollho R, Djouguela A, Rühmann A, Willborn KC and Harder D. DAVID-a translucent multi-wire transmission ionization chamber for in vivo verication of IMRT and conformal irradiation techniques. Phys Med Biol 2006; 51:1237-48.
6. [9] Poppe B, Looe H K, Chofor N, Rühmann A, Harder D and Willborn K. Clinical Performance of a Transmission Detector Array for the Permanent Supervision of IMRT Deliveries. Radiother. Oncol. 2010; 95:158-65
7. [10] Looe H K, Harder D, Rühmann A, Willborn K and Poppe B. Enhanced accuracy of the permanent surveillance of IMRT deliveries by iterative deconvolution of DAVID chamber signal proles Phys. Med. Biol. 2010; 55:3981-92
8. [11] Heukelom, S., el al.Wedge factor constituents of high-energy photon beams: Head and phantom scatter dose components Radiother. Oncol. 32: (1994) 66-3
9. 12] Jursinic, P. A. Changes in incident photon uence of 6 and 18 MV x rays caused by blocks and block trays Med Phys 26 (1999) 2092-8
10. [13] v. Klevens, H. Dependence of the tray transmission factor on collimator setting and sourcesurface distance Med. Phys. 27 (2000) 2117-3
11. [14] Sharma, S.C., el al., Recommendations for measurement of tray and wedge factors for high energy photons Med Phys 21 (1994) 573-5
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg47
Thank you for your attention
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg48
Additional Slides
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg
Description of the Elekta synergy DAVID160 system
50
David58
David160
WG Medical Radiation Physics, Pius-Hospital and Carl von Ossietzky University, Oldenburg51
• Φ constant with depth (small # interactions)
• Same # electrons set in motion in each square
• i.e., interactions per volume constant through target
• dose reaches a maximum at R (kerma constant with depth, equals absorbed dose beyond )
Number of electron tracks set in motion by photon interaction