Technical Information FLXA402 4 線式液分析計 HART 通信 TI 12A01F01-61JA 2018.11 初版(YK) 2019.07 2 版(YK) TI 12A01F01-61JA 目次 1. 概要 ............................................................................................................ 2 1.1 DD ファイルのダウンロードとインストール.................................................... 2 1.2 HART 通信の機能................................................................................................. 3 1.2.1 マルチドロップモード .................................................................................................4 1.2.2 設定の保護機能 ..............................................................................................................4 1.2.3 Device Variable Code ....................................................................................................5 1.2.4 Diagnostics Command (Command 48) .................................................................... 6 1.3 操作注意事項...................................................................................................... 10 2. DD メニュー構造(機器レビジョン 1、DD レビジョン 1) ........ 11 2.1 オンラインメニュー .......................................................................................... 11 2.1.1 process_variables_root_menu ............................................................................... 11 2.1.2 diagnostic_root_menu ............................................................................................. 16 2.1.3 maintenance_root_menu........................................................................................ 17 2.1.4 device_root_menu .................................................................................................... 23 2.2 オフライン ......................................................................................................... 35 改版履歴............................................................................................................ 37 「FLEXA」、「FLXA」、「SENCOM」、「FieldMate」は、横河電機の商標または登録商標です。 本文中に使われている会社名・商品名は、各社の登録商標または商標です。TM® マークは表示しておりません。

Upload: others

Post on 17-Sep-2020

51 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TechnicalInformation

FLXA402 4線式液分析計

HART 通信

TI 12A01F01-61JA2018.11 初版(YK)2019.07 2 版(YK)

TI 12A01F01-61JA

目次1. 概要............................................................................................................ 2

1.1 DDファイルのダウンロードとインストール.................................................... 21.2 HART 通信の機能................................................................................................. 3

1.2.1 マルチドロップモード.................................................................................................41.2.2 設定の保護機能..............................................................................................................41.2.3 DeviceVariableCode....................................................................................................51.2.4 DiagnosticsCommand(Command48)....................................................................6

1.3 操作注意事項...................................................................................................... 10

2. DDメニュー構造(機器レビジョン 1、DDレビジョン 1)........112.1 オンラインメニュー.......................................................................................... 11

2.1.1 process_variables_root_menu............................................................................... 112.1.2 diagnostic_root_menu............................................................................................. 162.1.3 maintenance_root_menu........................................................................................ 172.1.4 device_root_menu.................................................................................................... 23

2.2 オフライン......................................................................................................... 35

改版履歴............................................................................................................37

「FLEXA」、「FLXA」、「SENCOM」、「FieldMate」は、横河電機の商標または登録商標です。本文中に使われている会社名・商品名は、各社の登録商標または商標です。TM、® マークは表示しておりません。

Page 2: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

2 HART 通信

本書では、FLXA4024線式液分析計のHART通信について解説しています。パラメータの詳細については、FLXA4024線式液分析計の取扱説明書 変換器操作編(IM12A01A03-01JA)に記載されている内容を理解のうえ、HART通信を行ってください。

FLXA402 の最新版の電子マニュアルは、次のサイトからダウンロードできます。 http://www.yokogawa.co.jp/an/flxa402/download/

1. 概要HART 通信は、4-20mADC のアナログ信号に重畳されたHART 通信波形を使用し、稼動状態(オンライン)の FLXA402 と設定ツール(※)との相互通信を遠隔で行うものです。※: 設定ツールには、FieldMate、PRM、HHT携帯型通信ターミナルなどを用いることができます。注: HART 通信を行うツールに、FieldMate を使用する場合は、下記に示すVersion より新しいものをご使用

ください。【FieldMate】 FieldMateR3.04.00+DeviceFilesR3.09.00

HART は FIELDCOMMGROUP(https://www.fieldcommgroup.org/)の登録商標です。

1.1 DDファイルのダウンロードとインストールFLXA402 と設定ツール間でHART 通信を行うには、DD(DeviceDescription)ファイルを設定ツールにインストールする必要があります。DDファイルは、HART 通信の内容やFLXA402 固有のメニュー構成を含んだファイルです。FieldMate には、FieldMate 販売時点における最新のDDファイルが提供されています。最新のDDファイルは、以下のURLからダウンロードすることができます(※)。横河電機HP:http://www.yokogawa.com/an/download/an-dl-fieldbus-001en.htm※: URL は予告無く変更されることがあります。上記URLにアクセスできない場合は、当社各営業拠点また

はご購入の代理店にお問い合わせください。

■FieldMateを使用するにはFieldMate を使用する場合は、DeviceFiles のレビジョンに注意が必要です。

FLXA402とFieldMate、DDのレビジョン対応表HOUSINGASSY FieldMate

ソフトウェアレビジョン 機器レビジョン DeviceFilesレビジョン1.01.01 以降 1 R3.09 以降

2019.07.01-00No. TI 12A01F01-61JA 2nd Edition : July 2019(YK)All Rights Reserved. Copyright © 2018, Yokogawa Electric Corporation

Page 3: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 3

ソフトウェアレビジョンは、FLXA402 表示器のメイン画面もしくはホーム画面の変換器メニューボタン ををタップし、詳細画面( )をタップすると表示し、図 1の画面で確認することができます。

FLXA402ソフトウェアのレビジョン

機器レビジョン

図1 ソフトウェアレビジョンの表示画面

■DDファイルのインストールDDファイルのインストール方法については、お手持ちの設定ツール(FieldMate 等 ) の取扱説明書を参照してください。FieldMateR3.04.00+DeviceFilesR3.09.00 をご使用の場合は、DDのインストールは不要です。

■接続方法設定ツールの接続端子とmA1に設置した 250Ω以上の負荷抵抗の両端にHARTモデムを接続してください。計器室、機器端子箱、伝送ループ内のどの中継端子に対しても接続して使用することができます。

1.2 HART通信の機能FLXA402 の HART 通信は、エラー(4.7 節)、電流出力(4.3 節)、接点入力(4.2 節)、HART機能(4.6.2項)の設定が可能です。センサの設定情報や校正情報は参照のみ可能です。( )内は、変換器操作編(IM12A01F01-03JA)の参照項目です。

【メニュー構造と本体側画面との対応】 +–ConverterMenu ���本体の Slot1 の検出器情報を表示  +–HOLD ���手動HOLDが可能  +–WASH ���操作不可  +–Detail ���参照のみ可能  +–Trend ���参照不可  +–Other ���参照不可  +–Reset ���参照不可  +–Setting ���maOutput、maInput、HART 機能の設定変更が可能  +–Lang ���参照不可  +–Alarm ���参照可能 +–sensorMenu ���本体の Slot1 の検出器情報を表示  +–Detail ���参照のみ可能  +–Calibration ���実行不可  +–ResetWellness ���実行不可  +–Reset ���実行不可  +–Setting ���参照のみ可能

2019.07.01-00

Page 4: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

4 HART 通信

1.2.1 マルチドロップモードマルチドロップモードに設定した機器は、1本の通信伝送ライン上に複数のHART通信機器を接続することができます。マルチドロップ通信を有効にするには、機器アドレスを 1~ 63のいずれかに設定してください。

注記マルチドロップモードでHART 通信を行うには、HART 通信機器に設定する機器アドレスにはユニーク(固有)なポーリングアドレスを設定する必要があります。同じポーリングアドレスを 2台以上に設定した場合は、正常に通信できません。

マルチドロップモードの解除マルチドロップモードに設定された機器をマルチドロップモードから解除するには、ポーリングアドレスを 0に設定してください。

1.2.2 設定の保護機能スライドスイッチをONにすると、HART通信による設定ができなくなります。出荷時はOFFになっています。画面で「設定」のパスワードを設定した場合は、HARTも設定禁止状態にできます。変換器操作編(IM12A01F01-03JA)の 1.3 節を参照してください。

1

ON

2

前面扉を開いた状態

スライドスイッチ(HART 使用時のみ)リブートスイッチ

( 使用しないでください)

図2 FLXA402 各部の名称

注意スライドスイッチの1がHART用です。スライドスイッチの2はONにしないでください。

注意スライドスイッチの設定 /解除のときに、隣にあるスイッチ(押しボタン)を押さないよう、ご注意ください。

2019.07.01-00

Page 5: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 5

1.2.3 DeviceVariableCodeFLXA402 の DeviceVariableCode は、SVからQVの設定または Command9 で測定値をReadする際に使用します(下表参照)。Code Name PV SV TV QV Code Name PV SV TV QV0 AO1current*1 - X X X 25 Temp1-1 X X X X1 AI X X X X 26 pH1-1 X X X X2 DifferentialpH X X X X 27 ORP1-1 X X X X3 DifferentialORP X X X X 28 rH1-1 X X X X4 DifferentialConduct X X X X 29 SensormV11-1 - X X X5 DifferentialResist X X X X 30 SensormV21-1 - X X X6 DifferentialDO X X X X 31 SensormV31-1 - X X X7 AveragepH X X X X 32 Conduct11-1 X X X X8 AverageORP X X X X 33 Conduct21-1 X X X X9 AverageConduct X X X X 34 Resist11-1 X X X X10 AverageResist X X X X 35 Resist21-1 X X X X11 AverageDO X X X X 36 Concent11-1 X X X X12 Ratio X X X X 37 Concent21-1 X X X X13 Passage X X X X 38 USP1-1 - X X X14 Rejection X X X X 39 SensorOhms1-1 - X X X15 Deviation X X X X 40 Oxygen1-1 X X X X16 pHcalc X X X X 41 Sensorcurrent1-1 - X X X17 Script1 X X X X *18 Script2 X X X X *19 Script3 X X X X *20 Script4 X X X X *21 Script5 X X X X *22 Script6 X X X X *23 Script7 X X X X *24 Script8 X X X X *

42 Temp2-1 X X X X 59 Temp1-2 X X X X43 pH2-1 X X X X 60 pH1-2 X X X X44 ORP2-1 X X X X 61 ORP1-2 X X X X45 rH2-1 X X X X 62 rH1-2 X X X X46 SensormV12-1 - X X X 63 SensormV11-2 - X X X47 SensormV22-1 - X X X 64 SensormV21-2 - X X X48 SensormV32-1 - X X X 65 SensormV31-2 - X X X49 Conduct12-1 X X X X 66 Conduct11-2 X X X X50 Conduct22-1 X X X X 67 Conduct21-2 X X X X51 Resist12-1 X X X X 68 Resist11-2 X X X X52 Resist22-1 X X X X 69 Resist21-2 X X X X53 Concent12-1 X X X X 70 Concent11-2 X X X X54 Concent22-1 X X X X 71 Concent21-2 X X X X55 USP2-1 - X X X 72 USP1-2 - X X X56 SensorOhms2-1 - X X X 73 SensorOhms1-2 - X X X57 Oxygen2-1 X X X X 74 Oxygen1-2 X X X X58 Sensorcurrent2-1 - X X X 75 Sensorcurrent1-2 - X X X

76 Temp1-3 X X X X 93 Temp1-4 X X X X77 pH1-3 X X X X 94 pH1-4 X X X X78 ORP1-3 X X X X 95 ORP1-4 X X X X79 rH1-3 X X X X 96 rH1-4 X X X X80 SensormV11-3 - X X X 97 SensormV11-4 - X X X81 SensormV21-3 - X X X 98 SensormV21-4 - X X X82 SensormV31-3 - X X X 99 SensormV31-4 - X X X83 Conduct11-3 X X X X 100 Conduct11-4 X X X X84 Conduct21-3 X X X X 101 Conduct21-4 X X X X85 Resist11-3 X X X X 102 Resist11-4 X X X X86 Resist21-3 X X X X 103 Resist21-4 X X X X87 Concent11-3 X X X X 104 Concent11-4 X X X X88 Concent21-3 X X X X 105 Concent21-4 X X X X89 USP1-3 - X X X 106 USP1-4 - X X X90 SensorOhms1-3 - X X X 107 SensorOhms1-4 - X X X91 Oxygen1-3 X X X X 108 Oxygen1-4 X X X X92 Sensorcurrent1-3 - X X X 109 Sensorcurrent1-4 - X X X

X:設定可  -:設定不可*:現在は非数 (NaN) しか表示しません。

2019.07.01-00

Page 6: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

6 HART 通信

FLXA402 は、以下の単位コードを独自で拡張してます。Unit UnitsCode

rH 240S/m 241µS/m 242S/cm 243Ohm*m 244MOhm*m 245%SAT 247

1.2.4 DiagnosticsCommand(Command48)HARTStatusGroup CMD48

byte bit HARTAlarmName AlarmNumber NE107

DeviceStatus - 0 PVOutofLimit - S1 Non-PVOutofLimit - S2 LoopCurrentSaturated - S3 LoopCurrentFixed - N4 MoreStatusAvailable - N5 ColdStart - N6 ConfigurationChanged - N7 DeviceMalfunction - N

DeviceSpecificStatus0 0 0 Hardwarefailure 001 F1 Internalcom.error 002 F2 IOmod.param.readerror 003 F3 Com.mod.param.readerror 004 F4 CPUparam.readerror 005 F5 Reserved 006 N6 Reserved 007 N7 Reserved 008 N

DeviceSpecificStatus1 1 0 VGBcalc.error 009 F1 Ratiocalc.error 010 F2 Passagecalc.error 011 F3 Rejectioncalc.error 012 F4 Deviationcalc.error 013 F5 Scripterror 014 N6 Reserved 015 N7 Reserved 016 N

DeviceSpecificStatus2 2 0 mA1outputburnout 017 F1 mA2outputburnout 018 F2 mA3outputburnout 019 F3 mA4outputburnout 020 F4 mA1saturation 021 S5 mA2saturation 022 S6 mA3saturation 023 S7 mA4saturation 024 S

DeviceSpecificStatus3 3 0 AIexceedslimit 025 S1 AIoutofrange 026 F2 Expirytimeexceeded 027 M3 Reserved N4 Washresponsefailure 029 M5 Reserved N6 ConverterErrorSimulation 031 C7 Failsafeoccur 032 F

2019.07.01-00

Page 7: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 7

HARTStatusGroup CMD48byte bit HARTAlarmName Alarm

Number NE107

DeviceSpecificStatus4 4 0 SimulatedmAvalue1 033 C1 SimulatedmAvalue2 034 C2 SimulatedmAvalue3 035 C3 SimulatedmAvalue4 036 C4 OutputsinHOLD1 037 N5 OutputsinHOLD2 038 N6 OutputsinHOLD3 039 N7 OutputsinHOLD4 040 N

DeviceSpecificStatus5 5 0 Outputconfigurationerror 041 F1 Otherconfigurationerror 042 C2 Reserved 043 N3 Reserved 044 N4 Reserved 045 N5 Reserved 046 N6 Reserved 047 N7 Reserved 048 N

ExtendedDeviceStatus 6 0 MaintenanceRequired(NE107:M) - N1 DeviceVariableAlert*1 - S2 CriticalPoewFailure F3 Failure(NE107:F) - N4 OutofSpecification(NE107:S) - N5 FunctionCheck(NE107:C) - N6 Reserved N7 Reserved N

DeviceOperationgMode

7 0 Reserved N1 Reserved N2 Reserved N3 Reserved N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

StandardizedStatus0 8 0 DeviceVariableSimulationActive C1 Non-VolatileMemoryDefect F2 VolatileMemoryDefect F3 WatchdogResetExecuted F4 PowerSupplyConditionsOutofRange S5 EnvironmentalConditionsOutofRange S6 ElectronicDefect F7 DeviceConfigurationLocked - N

StandardizedStatus1 9 0 StatusSimulationActive - N1 DiscreteVariableSimulationActive - C2 EventNotificationOverflow N3 Reserved N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

AnalogChannelSaturated

10 0 AnalogChannel1 N1 AnalogChannel2 N2 AnalogChannel3 N3 AnalogChannel4 N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

2019.07.01-00

Page 8: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

8 HART 通信

HARTStatusGroup CMD48byte bit HARTAlarmName Alarm

Number NE107

StandardizedStatus2 11 0 Sub-DeviceListChanged N1 DuplicatemasterDetected. N2 Reserved N3 Reserved N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

StandardizedStatus2forWirelessHART

12 0 Reserved N1 Reserved N2 Reserved N3 Reserved N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

AnalogChannelFixed 13 0 AnalogChannel1 N1 AnalogChannel2 N2 AnalogChannel3 N3 AnalogChannel4 N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

DeviceSpecificStatus6 14 0 Sensor1-1devicefailure 101 F1 Sensor1-2devicefailure 201 F2 Sensor1-3devicefailure 301 F3 Sensor1-4devicefailure 401 F4 Sensor2-1devicefailure 501 F5 Reserved 601 N6 Reserved 701 N7 Reserved 801 N

DeviceSpecificStatus7 15 0 Sensor1-1configurationerror 102 F1 Sensor1-2configurationerror 202 F2 Sensor1-3configurationerror 302 F3 Sensor1-4configurationerror 402 F4 Sensor2-1configurationerror 502 F5 Reserved 602 N6 Reserved 702 N7 Reserved 802 N

DeviceSpecificStatus8 16 0 Temp.1-1exceedsuserlimit 103 S1 Temp.1-2exceedsuserlimit 203 S2 Temp.1-3exceedsuserlimit 303 S3 Temp.1-4exceedsuserlimit 403 S4 Temp.2-1exceedsuserlimit 503 S5 Reserved 603 N6 Reserved 703 N7 Reserved 803 N

DeviceSpecificStatus9 17 Process1-1exceedsuserlimit 104 S1 Process1-2exceedsuserlimit 204 S2 Process1-3exceedsuserlimit 304 S3 Process1-4exceedsuserlimit 404 S4 Process2-1exceedsuserlimit 504 S5 Reserved 604 N6 Reserved 704 N7 Reserved 804 N

2019.07.01-00

Page 9: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 9

HARTStatusGroup CMD48byte bit HARTAlarmName Alarm

Number NE107

DeviceSpecificStatus10 18 0 Sensor1-1measurementwarning 105 N1 Sensor1-2measurementwarning 205 N2 Sensor1-3measurementwarning 305 N3 Sensor1-4measurementwarning 405 N4 Sensor2-1measurementwarning 505 N5 Reserved 605 N6 Reserved 705 N7 Reserved 805 N

DeviceSpecificStatus11 19 0 Sensor1-1outofspec. 106 S1 Sensor1-2outofspec. 206 S2 Sensor1-3outofspec. 306 S3 Sensor1-4outofspec. 406 S4 Sensor2-1outofspec. 506 S5 Reserved 606 N6 Reserved 706 N7 Reserved 806 N

DeviceSpecificStatus12 20 0 Sensor1-1warning 107 N1 Sensor1-2warning 207 N2 Sensor1-3warning 307 N3 Sensor1-4warning 407 N4 Sensor2-1warning 507 N5 Reserved 607 N6 Reserved 707 N7 Reserved 807 N

DeviceSpecificStatus13 21 0 Sensor1-1disable 108 F1 Sensor1-2disable 208 F2 Sensor1-3disable 308 F3 Sensor1-4disable 408 F4 Sensor2-1disable 508 F5 Reserved 608 N6 Reserved 708 N7 Reserved 808 N

DeviceSpecificStatus14 22 0 Reserved N1 Reserved N2 Reserved N3 Reserved N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

DeviceSpecificStatus15 23 0 Reserved N1 Reserved N2 Reserved N3 Reserved N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

DeviceSpecificStatus16 24 0 Reserved N1 Reserved N2 Reserved N3 Reserved N4 Reserved N5 Reserved N6 Reserved N7 Reserved N

*: Reservedisfixed0.

2019.07.01-00

Page 10: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

10 HART 通信

1.3 操作注意事項本項は操作上注意していただきたいことを説明したものです。

●設定ツールとFLXA402での操作FLXA402 の設定変更は、HART 通信、画面設定、MODBUS 通信経由で同時変更することはできません。画面による設定変更中やMODBUS 通信による設定変更中は、HART 通信から設定を行うとエラーとなります。設定変更時は、電流出力設定を手動ホールド状態にして変更することを推奨します(変換器操作編(IM12A01F01-03JA)の 3.7 節参照)。

●パラメータの印刷やアップロード、ダウンロードなどの操作FLXA402 は大変多くのパラメータを保持しています。多くのパラメータにアクセスする動作では、数分かかることがあります。

●表示上の注意FLXA402 の設定を変更したら、メニュー表示を更新してください。

2019.07.01-00

Page 11: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 11

2. DDメニュー構造(機器レビジョン1、DDレビジョン1)

2.1 オンラインメニューprocess_variables_root_menudevice_root_menudiagnostic_root_menumaintenance_root_menu

2.1.1 process_variables_root_menuProcessvariablesrootmenu Dynamicvariables

PVPV%rngeSVTVQVPVisSVisTVisQVis

DevicevariablesCH1-1

SensortypePHsensor

pH1-1ORP1-1rH1-1Temp1-1SensormV11-1SensormV21-1SensormV31-1

SCsensorConduct11-1Conduct21-1Resist11-1Resist21-1Temp1-1Concent11-1Concent21-1USP1-1Sensorohms1-1

ISCsensorConduct11-1Conduct21-1Temp1-1Sensorohms1-1

DOsensorOxygen1-1Temp1-1Sensorcurrent1-1

2019.07.01-00

Page 12: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

12 HART 通信

CH1-2 �CH1-3 �CH1-4 �CH2-1 �

ConverterAO1currentAIDifferentialpHDifferentialORPDifferentialConductDifferentialResistDifferentialDOAveragepHAverageORPAverageConductAverageResistAverageDORatioPassageRejectionDeviationpHcalcScript1Script2Script3Script4Script5Script6Script7Script8

Dynamicvariablesstatus

PVqualityPVlimitSVqualitySVlimitTVqualityTVlimitQVqualityQVlimit

Devicevariablesstatus

CH1-1PHsensor

pH1-1qualitypH1-1limitORP1-1qualityORP1-1limitrH1-1qualityrH1-1limitTemp1-1qualityTemp1-1limitSensormV11-1qualitySensormV11-1limitSensormV21-1qualitySensormV21-1limitSensormV31-1qualitySensormV31-1limit

2019.07.01-00

Page 13: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 13

SCsensorConduct11-1qualityConduct11-1limitConduct21-1qualityConduct21-1limitResist11-1qualityResist11-1limitResist21-1qualityResist21-1limitTemp1-1qualityTemp1-1limitConcent11-1qualityConcent11-1limitConcent21-1qualityConcent21-1limitUSP1-1qualityUSP1-1limitSensorOhms1-1qualitySensorOhms1-1limit

ISCsensorConduct11-1qualityConduct11-1limitConduct21-1qualityConduct21-1limitTemp1-1qualityTemp1-1limitSensorOhms1-1qualitySensorOhms1-1limit

DOsensorOxygen1-1qualityOxygen1-1limitTemp1-1qualityTemp1-1limitSensorcurrent1-1qualitySensorcurrent1-1limit

CH1-2 �CH1-3 �CH1-4 �CH2-1 �

ConverterAO1currentqualityAO1currentlimitAIqualityAIlimitDifferentialpHqualityDifferentialpHlimitDifferentialORPqualityDifferentialORPlimitDifferentialConductqualityDifferentialConductlimitDifferentialResistquality

2019.07.01-00

Page 14: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

14 HART 通信

DifferentialResistlimitDifferentialDOqualityDifferentialDOlimitAveragepHqualityAveragepHlimitAverageORPqualityAverageORPlimitAverageConductqualityAverageConductlimitAverageResistqualityAverageResistlimitAverageDOqualityAverageDOlimitRatioqualityRatiolimitPassagequalityPassagelimitRejectionqualityRejectionlimitDeviationqualityDeviationlimitpHcalcqualitypHcalclimitScript1qualityScript1limitScript2qualityScript2limitScript3qualityScript3limitScript4qualityScript4limitScript5qualityScript5limitScript6qualityScript6limitScript7qualityScript7limitScript8qualityScript8limit

Viewinput/outputAO1currentAO2currentAO3currentAO4currentAIcurrentContactin/out

SensorvariablesCH1-1

PHsensorIMP11-1IMP21-1SensormV11-1SensormV21-1SensormV31-1

2019.07.01-00

Page 15: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 15

SCsensorSensorOhms1-1Polarization1-1

ISCsensorSensorOhms1-1

DOsensorSensorcurrent1-1Polarizationvoltage1-1AvailableKOH1-1Pressurelevel(process)

CH1-2 �CH1-3 �CH1-4 �CH2-1 �

Userselect Userselectvalues[selecteddevicevariavlename][selecteddevicevariavlename][selecteddevicevariavlename][selecteddevicevariavlename][selecteddevicevariavlename][selecteddevicevariavlename][selecteddevicevariavlename][selecteddevicevariavlename]

UserselectcodesUserselectcode1Userselectcode2Userselectcode3Userselectcode4Userselectcode5Userselectcode6Userselectcode7Userselectcode8

2019.07.01-00

Page 16: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

16 HART 通信

2.1.2 diagnostic_root_menuDiagnosticsrootmenu

DevicestatusDeviceStatusExtendedDeviceStatusStandardizedStatus0StandardizedStatus1DeviceSpecificStatus0DeviceSpecificStatus1DeviceSpecificStatus2DeviceSpecificStatus3DeviceSpecificStatus4DeviceSpecificStatus5DeviceSpecificStatus6DeviceSpecificStatus7DeviceSpecificStatus8DeviceSpecificStatus9DeviceSpecificStatus10DeviceSpecificStatus11DeviceSpecificStatus12DeviceSpecificStatus13ConfigchangedcountClearconfigchangedflag

DetaileddevicestatusCH1-1

PHsensorSCsensorISCsensorDOsensor

CH1-2 �CH1-3 �CH1-4 �CH2-1 �Converter

2019.07.01-00

Page 17: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 17

2.1.3 maintenance_root_menuMaintenancerootmenu

TestSimulationallclearTestautoreleasetimeLoopcurrent(AO1)testAO2testAO3testAO4testStatussimulation

DisplaySquawkAdjusttouchpanelLuminanceselect

ErrorconfigurationConverter

MeasurementalarmVGBcalc.errorRatiocalc.errorPassagecalc.errorRejectioncalc.errorDeviationcalc.errorScriptError1ScriptError2ScriptError3ScriptError4

I/OalarmFailsafeoccur

MaintenancestatusOutputsinHOLD1OutputsinHOLD2OutputsinHOLD3OutputsinHOLD4Washresponsefailure1Washresponsefailure2Washresponsefailure3Washresponsefailure4

SettingalarmmA1configurationerrormA2configurationerrormA3configurationerrormA4configurationerrorErrorinmAtable1ErrorinmAtable2ErrorinmAtable3ErrorinmAtable4Display1-1settingerrorDisplay1-2settingerrorDisplay1-3settingerrorDisplay1-4settingerrorDisplay2-1settingerrorDisplay2-2settingerrorDisplay2-3settingerrorDisplay2-4settingerror

2019.07.01-00

Page 18: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

18 HART 通信

HARTsettingerrorContactconfig.error1Contactconfig.error2Contactconfig.error3Contactconfig.error4

PHsensorSensorstatus

SensornotdetectTempelementnotdetectIDchipfailureLEdetectcalibrationdueimp1glasbreakdetectimp2glasbreakdetect

MeasurementwarningtemptoohightemptoolowpHtoohighpHtoolowORPtoohighORPtoolowrHtoohighrHtoolowimp(pH/ORP)toohighimp(pH/ORP)toolowimp(ref)toohighimp(ref)toolowpHtempcompwarningSSAtempout.Operatingspec

DevicestatusSENCOMComm.Error

SCsensorSensorstatus

SensornotdetectTempelementnotdetectIDchipfailurecalibrationdue

Measurementwarningtemptoohightemptoolowconductivitytoohighconductivitytoolowtempcomp1warningtempcomp2warningUSPlimitexceededUSPmarginexceededPolarizationdetectedmatrix1errormatrix2errorconctableerrorSSAtempout.Operatingspec

DevicestatusSENCOMComm.Error

2019.07.01-00

Page 19: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 19

ISCsensorSensorstatus

SensornotdetectTempelementnotdetectIDchipfailurecalibrationdue

Measurementwarningtemptoohightemptoolowconductivitytoohighconductivitytoolowtempcomp1warningtempcomp2warningmatrix1errormatrix2errorconctableerrorSSAtempout.Operatingspec

DevicestatusSENCOMComm.Error

DOsensorSensorstatus

SensornotdetectTempelementnotdetectIDchipfailuresensormembranefailurecalibrationdue

MeasurementwarningtemptoohightemptoolowDOtoohighDOtoolowSSAtempout.Operatingspec

DevicestatusSENCOMComm.Error

Condensedstatusmap

DevicestatusPrimaryVariableOutofLimitsNon-PrimaryVariableOutofLimitsLoopCurrentSaturatedLoopCurrentFixedMoreStatusAvailableColdStartConfigurationChangedDeviceMalfunction

ExtdevstatusMaintenancerequiredDevicevariablealert

2019.07.01-00

Page 20: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

20 HART 通信

FailureOutofSpecificationFunctionCheck

Devicediagnosticstatus0

Deviceconfigurationlocked

Devicediagnosticstatus1

StatussimulationactiveDiscreteVariableSimulationactive

DeviceSpecificStatus0

HardwarefailureInternalcom.errorIOmod.param.readerrorCom.mod.param.readerrorCPUparam.readerror

DeviceSpecificStatus1

VGBcalc.errorRatiocalc.errorPassagecalc.errorRejectioncalc.errorDeviationcalc.errorScripterror

DeviceSpecificStatus2

mA1outputburnoutmA2outputburnoutmA3outputburnoutmA4outputburnoutmA1saturationmA2saturationmA3saturationmA4saturation

DeviceSpecificStatus3

AIexceedslimitAIoutofrangeExpirytimeexceededWashresponsefailureConverterErrorSimulationFailsafeoccur

DeviceSpecificStatus4

SimulatedmAvalue1SimulatedmAvalue2SimulatedmAvalue3SimulatedmAvalue4OutputsinHOLD1

2019.07.01-00

Page 21: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 21

OutputsinHOLD2OutputsinHOLD3OutputsinHOLD4

DeviceSpecificStatus5

OutputconfigurationerrorOtherconfigurationerror

DeviceSpecificStatus6

Sensor1-1devicefailureSensor1-2devicefailureSensor1-3devicefailureSensor1-4devicefailureSensor2-1devicefailureSensor2-2devicefailureSensor2-3devicefailureSensor2-4devicefailure

DeviceSpecificStatus7

Sensor1-1configurationerrorSensor1-2configurationerrorSensor1-3configurationerrorSensor1-4configurationerrorSensor2-1configurationerrorSensor2-2configurationerrorSensor2-3configurationerrorSensor2-4configurationerror

DeviceSpecificStatus8

Temp.1-1exceedsuserlimitTemp.1-2exceedsuserlimitTemp.1-3exceedsuserlimitTemp.1-4exceedsuserlimitTemp.2-1exceedsuserlimitTemp.2-2exceedsuserlimitTemp.2-3exceedsuserlimitTemp.2-4exceedsuserlimit

DeviceSpecificStatus9

Process1-1exceedsuserlimit

2019.07.01-00

Page 22: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

22 HART 通信

Process1-2exceedsuserlimitProcess1-3exceedsuserlimitProcess1-4exceedsuserlimitProcess2-1exceedsuserlimitProcess2-2exceedsuserlimitProcess2-3exceedsuserlimitProcess2-4exceedsuserlimit

DeviceSpecificStatus10

Sensor1-1measurementwarningSensor1-2measurementwarningSensor1-3measurementwarningSensor1-4measurementwarningSensor2-1measurementwarningSensor2-2measurementwarningSensor2-3measurementwarningSensor2-4measurementwarning

DeviceSpecificStatus11

Sensor1-1outofspec.Sensor1-2outofspec.Sensor1-3outofspec.Sensor1-4outofspec.Sensor2-1outofspec.Sensor2-2outofspec.Sensor2-3outofspec.Sensor2-4outofspec.

DeviceSpecificStatus12

Sensor1-1warningSensor1-2warningSensor1-3warningSensor1-4warningSensor2-1warningSensor2-2warningSensor2-3warningSensor2-4warning

DeviceSpecificStatus13

Sensor1-1disableSensor1-2disableSensor1-3disableSensor1-4disableSensor2-1disableSensor2-2disable

2019.07.01-00

Page 23: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 23

Sensor2-3disableSensor2-4disable

Resetcondensedstatusmap

Devicereset

2.1.4 device_root_menuDevicerootmenu

BasicsetupAnalogoutput

AO1isAO1targetAO1parameter

AO1linearlimitvalue

AO1LRVAO1URVAO1subLRVAO1subURV

AO1tablelimitvalue

AO1LRVAO1URV

AO2isAO2targetAO2parameter

AO2linearlimitvalue

AO2LRVAO2URVAO2subLRVAO2subURV

AO2tablelimitvalue

AO2LRVAO2URV

AO3isAO3targetAO3parameter

AO3linearlimitvalue

AO3LRVAO3URVAO3subLRVAO3subURV

AO3tablelimitvalue

AO3LRVAO3URV

2019.07.01-00

Page 24: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

24 HART 通信

AO4isAO4targetAO4parameter

AO4linearlimitvalue

AO4LRVAO4URVAO4subLRVAO4subURV

AO4tablelimitvalue

AO4LRVAO4URV

HARTbasicsetup

PVisSVisTVisQVis

TagsettingTagLongtag

DetailedsetupAOholdforsetupAnalogoutput/input

Analogoutput1AO1is

AO1targetAO1parameter

AO1modeAO1functionselectAO1burnoutAO1dampingtimeAO1simulationvalueAO1holdsetting

AO1holdLast/FixedselectAO1holdfixedvalueAO1holdduringcalibration

AO1linearlimitvalue

AO1LRVAO1URVAO1subLRVAO1subURV

AO1tablelimitvalue

2019.07.01-00

Page 25: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 25

AO1LRVAO1URV

Analogoutput2 �Analogoutput3 �Analogoutput4 �Analoginput

AItypeAIdampingtimeAIupperLimitAIlowerlimitAItemperature

AItempunitAITempLRVAITempURV

AIpressureAIpressunitAIPressLRVAIPressURVPressurecomp.(DO70G)

ContactinputContactinput1

DI1typeDI1washtargetDI1changerangetarget

Contactinput2DI2typeDI2washtargetDI2changerangetarget

DisplaysetupAutoreturntimeselectLuminanceselectBacklightautoofftimeselectSquawksetting

Deviceinformation

Date/TimeCurrentdateCurrenttime

SerialNo.SoftwarerevInternalserialNo.

HousingassymAoutputmoduleCommunicationmoduleAnalogsensormodule1Analogsensormodule2

2019.07.01-00

Page 26: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

26 HART 通信

Sensorinformation * 続きは別に記載

HARTdetailedsetup

HARToutputPVisSVisTVisQVisLoopcurrentmodePolladdr

HARTidentification

TagLongtagNumreqpreamsNumresppreamsDescriptorMessageDateFinalasemblynumDistributorModelWriteprotectSTXCountACKCount

HARTsubunitsetup

SubunitselectSensor1-1subunitenableSensor1-2subunitenableSensor1-3subunitenableSensor1-4subunitenableSensor2-1subunitenableSensor2-2subunitenableSensor2-3subunitenableSensor2-4subunitenable

HARTrevisionsFlddevrevDevidUniversalrevSoftwarerevHardwarerev

* Sensorinformationの続き

Sensorinformation

CH1-1SensortypeSerialNo.PHsensor

2019.07.01-00

Page 27: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 27

Sensorwellness

ProgresstimeHeatcycleZeroSlopeImpedance1Impedance2

Predictmaintenance

ProjectedmaintenancetimeProjectedmaintenancestatusProjectedreplacementtimeProjectedreplacementstatus

Settinginformation

ConfiguresensorSensortypeTemp.elementModbusaddress

MeasuresettingTemperaturesettings

Unit

Temp.compensation

CompensationManualtemp.Referencetemp.

ProcessTemp.Compensation

Method(pH)Tempcoef(TCpH)Method(ORP)Tempcoef1(TCORP)Tempcoef2(TCORP)Impedancemeasure

High/Lowalarmsetting

Temp.highlimitTemp.lowlimitpHhighlimitpHlowlimitORPhighlimitORPlowlimitrHhighlimitrHlowlimit

Calibrationsettings

Tempoffset

2019.07.01-00

Page 28: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

28 HART 通信

StabilizationtimeCalibr.IntervalCal.setpH

UnitZerounitSlopeunit

Limitsandtiming

zerohighlimitzerolowlimitslopehighlimitslopelowlimitStepRange

Buffers(selectset)

SelectBuffer

Zero/Slope/ITPZeroSlopeITPZero2Slope2

Cal.setORP/rHLimitsandtiming

zerohighlimitzerolowlimitslopehighlimitslopelowlimitStepRange(ORP)StepRange(rH)

zero/slopeZeroSlope

zero/slope2ZeroSlope

WellnesssettingsImpedance1

Impedance1HighlimitLowlimit

Impedance2Impedance2

2019.07.01-00

Page 29: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 29

HighlimitLowlimit

Sensordiag.settings

Impedance1FINEImpedance2FINEProgresstimeBADLimit

DefineheatcycleHeatcycleBADLimitHeatcycletemperatureHeatcycletime

DefineSENCOMstatus

Sterilizedtemp.SterilizedtimeHightemp.1Hightemp.2HighpHvalueLowpHvalue

SCsensorSensorwellness

ProgresstimeHeatcycleCellconstantPolarization

Predictmaintenance

ProjectedcalibrationtimeProjectedcalibrationstatus

Settinginformation

ConfiguresensorSensortypeMeasuringunitTemperaturesettings

Temp.element

Othersc.c.(factory)c.c.(adjusted)Modbusaddress

MeasuresettingSC1sel./comp.SC2sel./comp.

2019.07.01-00

Page 30: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

30 HART 通信

Concent1sel./comp.Temperaturesettings

Unit

Temp.compensation

CompensationManualtemp.Referencetemp

TempcoefTempcoef1Tempcoef2

MatrixSelectmatrix1Selectmatrix2

High/Lowalarmsetting

TemperatureHighLimitTemperatureLowLimitConductivityHighLimitConductivityLowLimitResistanceHighLimitResistanceLowLimitUSPsafetymargin

Calibrationsettings

TempoffsetLimits

Aircalibrationc.c.highc.c.low

TimingStepRangeStabilizationtimeCalibr.interval

WellnesssettingsSensordiag.settings

ProgresstimeBADLimit

Defineheatcycle

HeatcycleBADLimitHeatcycletemperature

2019.07.01-00

Page 31: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 31

Heatcycletime

OthersPolarizationHighLimit

DefineSENCOMstatus

Sterilizedtemp.SterilizedtimeHightemp.1Hightemp.2

ISCsensorSensorwellness

ProgresstimeHeatcycleCellconstant

Predictmaintenance

ProjectedcalibrationtimeProjectedcalibrationstatus

Settinginformation

Configuresensor

MeasuringunitTemperaturesettings

Temp.element

Othersc.c.(factory)c.c.(adjusted)Modbusaddress

MeasuresettingSC1sel./comp.SC2sel./comp.Concent1sel./comp.Temperaturesettings

Unit

Temp.compensation

CompensationManualtemp.Referencetemp

TempcoefTempcoef1Tempcoef2

Matrix

2019.07.01-00

Page 32: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

32 HART 通信

Selectmatrix1Selectmatrix2

High/Lowalarmsetting

TemperatureHighLimitTemperatureLowLimitConductivityHighLimitConductivityLowLimit

Calibrationsettings

TempoffsetLimits

Aircalibrationc.c.highc.c.low

TimingStepRangeStabilizationtimeCalibr.interval

WellnesssettingsSensordiag.settings

ProgresstimeBADLimit

DefineheatcycleHeatcycleBADLimitHeatcycletemperatureHeatcycletime

DefineSENCOMstatus

Sterilizedtemp.SterilizedtimeHightemp.1Hightemp.2

DOsensorSensorwellness

ProgresstimeHeatcycleSlope

Settinginformation

ConfiguresensorSensortypeGalvanic

SensorsensitivitySensorsensitivity(Userdefined)

2019.07.01-00

Page 33: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 33

PolarographicSensorsensitivityPolarizationVoltage

OthersTemp.elementModbusaddress

MeasuresettingDOsettings

Unit

Temperaturesettings

Unit

Temp.compensation

CompensationManualtemp.

Salinitycomp.CompensationSalinity

Pressurecomp.(Measure)

CompensationPressurelevel(process)Pressureunit

High/Lowalarmsetting

Temp.highlimitTemp.lowlimitDO(mgl)highlimitDO(mgl)lowlimitDO(ppm)highlimitDO(ppm)lowlimitDO(ppb)highlimitDO(ppb)lowlimitDO(%SAT)highlimitDO(%SAT)lowlimit

CalibrationsettingsTempoffsetLimitsandtiming(Galvanic)

zerohighlimitzerolowlimitslopehighlimitslopelowlimit

2019.07.01-00

Page 34: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

34 HART 通信

Limitsandtiming(Polarographic)

zerohighlimitzerolowlimitslopehighlimitslopelowlimitStabilizationtime

StepRange(mgl)zeropointspanpoint

StepRange(ppm)zeropointspanpoint

StepRange(ppb)zeropointspanpoint

StepRange(%SAT)

zeropointspanpointCalibr.interval

zero/slope(Galvanic)

ZeroCurrentSlope

zero/slope(Polarographic)

ZeroCurrentSlope

PressureComp.(Cal.)

Pressurelevel(calibration)

WellnesssettingsSensordiag.settings

ProgresstimeBADLimit

DefineheatcycleHeatcycleBADLimitHeatcycletemperatureHeatcycletime

CheckKOHresidue

Sensortype

DefineSENCOMstatus

Sterilizedtemp.SterilizedtimeHightemp.1

2019.07.01-00

Page 35: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

HART 通信 35

Hightemp.2

CH1-2 �CH1-3 �CH1-4 �CH2-1 �

2.2 オフラインoffline_root_menu Tag

LongtagDescriptorMessageDateFinalasmblynumPolladdrLoopcurrentmodeNumresppreamsTestautoreleasetime

SVisTVisQVis

AO1targetAO1parameterAO1modeAO1functionselectAO1burnoutAO1dampingtimeAO1simulationvalueAO1holdLast/FixedselectAO1holdfixedvalueAO1holdduringcalibration

AO2targetAO2parameterAO2modeAO2functionselectAO2burnoutAO2dampingtimeAO2simulationvalueAO2holdLast/FixedselectAO2holdfixedvalueAO2holdduringcalibration

AO3targetAO3parameterAO3modeAO3functionselectAO3burnoutAO3dampingtimeAO3simulationvalueAO3holdLast/FixedselectAO3holdfixedvalueAO3holdduringcalibration

AO4target

2019.07.01-00

Page 36: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

36 HART 通信

AO4parameterAO4modeAO4functionselectAO4burnoutAO4dampingtimeAO4simulationvalueAO4holdLast/FixedselectAO4holdfixedvalueAO4holdduringcalibration

AItypeAIdampingtimeAIupperLimitAIlowerlimitAITemp.LRVAITemp.URVAIPress.LRVAIPress.URVPressurecomp.(DO70G)

DI1typeDI1washtargetDI1changerangetargetDI2typeDI2washtargetDI2changerangetarget

Subunitselect

AutoreturntimeselectLuminanceselectBacklightautoofftimeselect

2019.07.01-00

Page 37: FLXA402 4 線式液分析計 HART通信47 Sensor mV2 2-1 - X X X 64 Sensor mV2 1-2 - X X X 48 Sensor mV3 2-1 - X X X 65 Sensor mV3 1-2 - X X X 49 Conduct12-1 X X X X 66 Conduct11-2

TI 12A01F01-61JA

37

改版履歴資料名称 : FLXA402 4線式液分析計 HART 通信

資料番号 : TI12A01F01-61JA

2019年7月/2版 変更(P.5、6)2018年11月/初版 新規作成