Radiocirugía Estereotáxica usando VMAT
(Arcoterapia Volumétrica Modulada)
Lic. Andrés GonzalesFísico Médico
![Page 2: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/2.jpg)
Av. José Gálvez Barrenechea 1044, San Isidro
![Page 3: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/3.jpg)
ACELERADOR LINEAL TRILOGY (VARIAN)
ALIADA CONTRA EL CÁNCER
![Page 4: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/4.jpg)
1968Radiación de Arcos
con campos
conformados pero no
modulados
1995Radiación de Arcos con
campos de intensidad
modulada
2002Radiación de Arcos en
helix
(Helical Tomotherapy)
2008Radiación de un Arco.
VMAT - RapidArc
Phys. Med. Biol. 40 (1995) 1435-
1449. Printed in the UK
First Treatment - University
of Wisconsin
Madison, Wisconsin
2002
2007 - RapidArc unveiled at
ASTRO - Los Angeles, CA.
2008 - Became clinically
available in USA – University of
Alabama.
2009/January – Became
clinically available at UC-Irvine.
2013 Julio – VMAT
RapidArc en el Perú -
ALIADA CONTRA EL
CÁNCER
Evolucion de la Tecnologia de RadiaciónTerapeutica Usando Arcos
![Page 5: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/5.jpg)
Proceso de SRS Radiocirugía Estereotáxica
TEM Simulación y
Máscara
Definición de
Volúmenes
Planificación Física
Control de Calidad
Tratamiento
Indispensable
![Page 6: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/6.jpg)
Acelerador Lineal con conos – Experiencia en Perú
• Alta dosis(16Gy – 24Gy)
• Radiocirugía Estereotáxica(única sesión de tratamiento)
• Volúmenes pequeños.
• Conos.• Marco
estereotáxico.
![Page 7: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/7.jpg)
UC Irvine
Trilogy en el UC Irvine
![Page 8: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/8.jpg)
Vision RT
Mesa 6° de libertad
TrueBeamUC Irvine
![Page 9: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/9.jpg)
Hoy en día, se reemplaza el uso del Marco Estereotáxico
![Page 10: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/10.jpg)
Metástasis Cerebrales
1. Técnica No-isocéntrica.
2. Se trata varias lesiones
a la vez.
3. Menor tiempo de
tratamiento.
4. Diferentes dosis a los
blancos.
5. No hay límite en el
tamaño de las lesiones.
6. Varía: Tasa de dosis /
velocidad de MLC /
Velocidad del gantry.
7. No se necesita marco
estereotáxico, mayor
comodidad para el
paciente.
![Page 11: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/11.jpg)
Planificación Física - Eclipse V.11
![Page 12: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/12.jpg)
TEM Simulación
![Page 13: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/13.jpg)
Planificación Física - Eclipse V.11
16Gy V > 3cc / 18Gy 1cc < V < 3cc / 20Gy V < 1cc
![Page 14: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/14.jpg)
Planificación Física - Eclipse V.11
![Page 15: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/15.jpg)
![Page 16: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/16.jpg)
Control de calidad con ArcCHECK
![Page 17: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/17.jpg)
![Page 18: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/18.jpg)
![Page 19: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/19.jpg)
Control de Calidad con Películas Radiocrómicas
Films GAFCHROMIC EBT3
![Page 20: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/20.jpg)
Control de calidad con TLD
![Page 21: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/21.jpg)
Control de calidad - Irradiación Fantoma Antropomófico
Tubo RX
Detector
![Page 22: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/22.jpg)
Posicionamiento con Cone Beam CT (CBCT)
CBCT
CT
TLD
![Page 23: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/23.jpg)
Vision RT
Varian
![Page 24: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/24.jpg)
Irradiación de Películas Dosimétricas EBT3
![Page 25: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/25.jpg)
![Page 26: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/26.jpg)
![Page 27: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/27.jpg)
![Page 28: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/28.jpg)
![Page 29: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/29.jpg)
![Page 30: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/30.jpg)
![Page 31: SRS Radiocirugia Estereotaxica usando VMAT _ ECI inv 2015](https://reader034.vdocuments.co/reader034/viewer/2022052414/55d11989bb61ebdf5d8b4754/html5/thumbnails/31.jpg)
1. La placa radiocrómica tiene mejor resolución
espacial que los equipos con arreglos de diodos.
1. VMAT-RapidArc es una Técnica No-isocéntrica.
2. Se trata varias lesiones a la vez.
3. Menor tiempo de tratamiento.
4. Diferentes dosis prescrita a los blancos.
5. No hay límite en el tamaño de las lesiones.
6. Varía: Tasa de dosis / velocidad de MLC / Velocidad del gantry.
7. Monitoreo del paciente en tiempo real (Vision RT).
8. No se necesita marco estereotáxico, mayor comodidad para el paciente.
9. Tratamiento ambulatorio.
10.Mejor calidad de vida - Protección del sistema Límbico: Es importante
evaluar los trastornos de las funciones superiores, memoria, atención
(Radioterapia Holocraneal).
11.Debe haber contante control de calidad al LINAC y dosimetría del equipo
por una entidad externa.
12. Entrenamiento del personal.
Conclusiones