Stability of polymersomes containing 111In using Perturbed Angular Correlation spectroscopy
Anca Tacu
Supervisor: Peter Bode
2Challenge the future
1. Introduction
2. Data Interpretation
3. Results & Discussion
4. Conclusion
5. Future Work
3Challenge the future
Cancer
Tumour tissueIncreased permeability of blood vessels
Dysfunctional lymphatic drainage
Accumulation nanocarriers
7Challenge the future
Loading efficiencylpmoltC
In/5.0)( 0111
lpmoltCCd
/86.1)( 0111
Efficiency of loading dependent on:• age of stock solution • type of buffer
8Challenge the future
PACMeasures correlation between the directions of emission of 2 consecutive γ-rays
Directional probability of emission dependent on:•orientation of the nuclear spin axis •surrounding electrical field
9Challenge the future
Stop,θ
Clock,t1
2
θ
Start
Differentiate between different surrounding media of the radionuclide studied
PAC
10Challenge the future
PAC
180 9022
22 22 180 90
12
N NG
A Q N N
180 9022
22 22 180 90
12
N NG
A Q N N
22 0.5G
Highly perturbed 0 ≤ G22 ≤ 1 No perturbation
11Challenge the future
PAC
Parameters:• Relaxation time• The skn(η) coefficients• The precession frequencies: ωkn(η)
),cos()(3
0
tsetG knn
knt
kkk
13Challenge the future
Results & Discussion3 sets of measurements:
• 111InCl3 crystalized and in solution at pH 2
• 111In-DTPA and 111In-Tropolone complexes
• 111In loaded polymersomes
14Challenge the future
Results & DiscussionSpecifications sample
Activity sample
Time since Calibration
Experimental Reference
Crystalized InCl3 300 kBq 10 d -0.15 0.14-0.02
Liquid 111InCl3 300 kBq 11 d 1.48 0.98Liquid 111InCl3 (liquid nitrogen)
0.08
22G
180 9022
22 22 180 90
12
N NGA Q N N
0 ≤ G22 ≤ 1A22 = -0.0870Q22 = 0.8066N180-N90<0
22G
15Challenge the future
Results & DiscussionSpecifications sample Activity
sampleTime since calibration
111In-DTPA in HEPES 130 kBq 17 d 0.31111In-DTPA in HEPES (liquid nitrogen) -0.57 111In-DTPA in PBS 0.57 111In-DTPA in PBS (liquid nitrogen) -1.93111In-Tropolone in HEPES 420 kBq 15 d 0.04 111In-Tropolone in PBS 0.23 111In-Tropolone in PBS (liquid nitrogen) -1.23
22G
Issue with the detectorOR
Correlation times of molecules too small to be detected by the PAC
16Challenge the future
Stability vs. Instability180 90
2222 22 180 90
12
N NGA Q N N
A22 = -0.0870Q22 = 0.8066
N180-N90<0
Assumption 180o detector stable: N180=90 000Case 1:
N90=60 000N90=100 000
Case 2:
22 0.67G 22 2.03G
17Challenge the future
Results & DiscussionSpecifications sample Activity
sampleTime since calibration
200 nm polymersomes in PBS
1st day 170 kBq 1 d 0.39 2nd day 130 kBq 2 d 0.26
200 nm polymersomes in HEPES
1st day 140 kBq 9 d 0.042nd day 110 kBq 10 d 0.17
400 nm polymersomes in HEPES
2nd day 140 kBq 10 d 0.12
800 nm polymersomes in HEPES
2nd day 170 kBq 10 d -0.05
22G
Size of polymersome seems not to have any influencePerturbation in HEPES polymersome seems to be higher than in PBS onesDoes the age of stock solution matter?
18Challenge the future
Conclusion• PAC has potential
• The system seems to have technical problems
• Winfit was not a reliable software
• No final conclusions could be drawn- more
measurements should be made
19Challenge the future
Future work1. Investigate:
• the influence of In3+
• the effect of Cd2+
• life-time of polymersomes upon systemic administration (in vivo & in vitro)
2. Was it just a problem with the detectors?
3. Purchase/ Implement a program for the data fitting