aqms presentation thermo

107
The world leader in serving science Ambient Air Quality Monitoring Presented By Naveen kumarService Engineer (EID) NOIDA

Upload: naveen-kumar

Post on 16-Jan-2017

859 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: AQMS PRESENTATION THERMO

The world leader in serving science

Ambient Air Quality

Monitoring

Presented By

Naveen kumar–Service Engineer (EID)

NOIDA

Page 2: AQMS PRESENTATION THERMO

2

POLLUTANTS EVERY WHERE - NEW POLLUTANTS - NEW STANDARDS - NEW MEASURING TECHNOLOGIES

POLLUTION …

Page 3: AQMS PRESENTATION THERMO

3

Are you interested in finding out about Air Quality?

Do you have questions in relation to Air Quality?

Look no further.

There is a lot of information, about Air Quality, and how we handles

Air Quality.

Air Quality

Page 4: AQMS PRESENTATION THERMO

4

Air pollution

Air pollution comes from many different sources: stationary sources such

as factories, power plants, and smelters and smaller sources such as dry

cleaners and degreasing operations; mobile sources such as cars, buses,

planes, trucks, and trains; and naturally occurring sources such as

windblown dust all contribute to air pollution.

Air Quality can be affected in many ways by the pollution emitted from

these sources. These pollution sources can also emit a wide variety of

pollutants.

The EPA has these pollutants classified as the six principal

pollutants (or criteria pollutants-as they are also known). These pollutants

are monitored by the EPA, as well as national, state and local

organizations.

Air Quality

Page 5: AQMS PRESENTATION THERMO

5

Finding out if the Air we Breathe is Clean

The six criteria pollutants addressed in the NAAQS are Carbon

Monoxide, Nitrogen Dioxide, Lead, Ozone (or smog), Particulate

Matter, and Sulfur Dioxide.

If the levels of these pollutants are higher than what is considered

acceptable by EPA, then the area in which the level is too high is

called a no attainment area.

The Air Pollution Monitoring program monitors all of the six criteria pollutants.

Measurements are taken to assess areas where there may be a problem, and

to monitor areas that already have problems. The goal of this program is to

control areas where problems exist and to try to keep other areas from

becoming problem air pollution areas.

POLLUTION

Page 6: AQMS PRESENTATION THERMO

6

What Are the Six Common Air Pollutants? O3

SO2

NOX

CO

SPM

LEAD

POLLUTION

Page 7: AQMS PRESENTATION THERMO

7

NAAQS - US EPA

The Clean Air Act, which was last amended in 1990, requires EPA

to set National Ambient Air Quality Standards for pollutants

considered harmful to public health and the environment.

The Clean Air Act established two types of national air quality

standards.

• Primary standards set limits to protect public health, including the health of

"sensitive" populations such as asthmatics, children, and the elderly.

• Secondary standards set limits to protect public welfare, including protection

against decreased visibility, damage to animals, crops, vegetation, and

buildings.

Page 8: AQMS PRESENTATION THERMO

8

… NAAQS - US EPA

The EPA Office of Air Quality Planning and Standards

(OAQPS) has set National Ambient Air Quality Standards for

six principal pollutants, which are called "criteria" pollutants:

PM10 & PM2.5, SO2, NO2, CO, O3, Lead

Units of measure for the standards are parts per million (ppm)

by volume, milligrams per cubic meter of air (mg/m3 ), and

micrograms per cubic meter of air (µg/m3 )

Page 9: AQMS PRESENTATION THERMO

9

International standards and guidelines (WHO-US EPA)

(1) : PM10 (<10 µm), (2) : PM 2.5 (<2.5 µm)

(3) : SP + SO2 : « black fumes » method (5 µm)

POLLUTANTS WHO(World Health Organization)

US EPA – NAAQS(National Ambient Air Quality Standard)

PARTICULATEMATTER

(PM) µg/m3

125 per 24h (3)50 per year (3)

65 per 24h (2)150 per 24h (1)15 per year (2)50 per year (1)

SULFUR DIOXIDE(SO2) µg/m

3

500 per 10-15 minutes125 per 24h50 per year

1300 per 3h (0.5 ppm)365 per 24h (0.14 ppm)80 per year (0.03 ppm)

NITROGEN DIOXIDE(NO2) µg/m3

200 per 1h40 per year

100 per year (0.053 ppm)

CARBON MONOXIDE(CO) mg/m3

100 per 10-15 minutes30 per 1h10 per 8h

40 per 1h (35 ppm)10 per 8h (9 ppm)

OZONE(O3) µg/m

3

120 per 8h 235 per 1h (0.12 ppm)

LEAD(Pb) µg/m

3

0.5 per year 1.5 per quarter

Page 10: AQMS PRESENTATION THERMO

10

Welcome to Thermo Fisher Scientific

The world leader in serving science

We are the leading provider of analytical instruments,

equipment, reagents and consumables, software and services

for research, analysis, discovery and diagnostics

Page 11: AQMS PRESENTATION THERMO

11

We enable our customers to make the world healthier, cleaner and safer

The World Leader in Serving Science

Scale

$10 billion in revenues

30,000 employees in 38 countries

350,000 customers

150 countries served

Capabilities

Complete portfolio

World-class technologies

Commercial and service strength

Experience 150 years of combined experience

Brand Equity Pre-eminent brands

Page 12: AQMS PRESENTATION THERMO

12

Page 13: AQMS PRESENTATION THERMO

13

Page 14: AQMS PRESENTATION THERMO

14

iSeries Platform – Design Overview

Page 15: AQMS PRESENTATION THERMO

15

Introduction

Page 16: AQMS PRESENTATION THERMO

16

iSeries Common Electronics - User Interface

Soft key prompts

One Button Programming

Page 17: AQMS PRESENTATION THERMO

17

iSeries Common Electronics - User Interface

Run

Menu

Title Bar • Shows what’s being displayed

Measurements • 1-5 lines

• Measurement descriptor

• Measurement value

• Units

Status bar • Measurement status

• Time

• Alarm status

Menu data • Menus, data display, entry

Page 18: AQMS PRESENTATION THERMO

18

iSeries New Features

Page 19: AQMS PRESENTATION THERMO

19

iSeries Platform – Common Electronics and Modular Measurement System

Common Electronics

Section

Measurement

System Module

Page 20: AQMS PRESENTATION THERMO

20

iSeries Common Electronics

Motherboard/Processor Board

Digital Output Board

Universal Switching

Power Supply

Graphic Display

Keypad

Front Panel

Connector Board

AC Power

Switch

Page 21: AQMS PRESENTATION THERMO

21

iSeries Common Electronics – Front View

Transformer for 220/240

VAC or 100 VAC

Operation

Page 22: AQMS PRESENTATION THERMO

22

iSeries Common Electronics – Rear View

RS485 Expansion Connector

(External Converter Boxes)

RS232/485 Serial Ports

Ethernet Port

Digital Output (Relay) Port

Common I/O Connector Expansion I/O Connector

Measurement System Connections

AC Power Connector

and Fuse

Page 23: AQMS PRESENTATION THERMO

23

iSeries New Features

FEATURES BENEFITS

Improved Layout Easier Component Accessibility

Product Commonality Fewer Parts

Shorter Lead Times

Quick Learning Curve

Eases Spare Parts Inventory

Ethernet Connectivity Efficient Remote Access

Enhanced User Interface Display Screen Customization

One Button Programming

Flash Memory Increased Data Storage

Web Based Firmware Upgrades

Page 24: AQMS PRESENTATION THERMO

24

iSeries Key Enhancements

Enhancement Details

Easier Maintenance Fewer PCBAs (33% reduction), case separation, fold-down

divider panel, component layout – better access, subassembly

captive screws, elimination of hardware based pot adjustments

Greater Product

Commonality

Identical motherboard, digital output board, I/O expansion board

and front panel board – eases spare parts inventory

Flash Memory Increased data storage (56K g 4mb)

Efficient firmware upgrades

Ethernet Connectivity Efficient analyzer remote access

User Interface Large graphics display and programmable soft keys

I/O Functionality cSeries: 8 digital inputs; 8 digital outputs; 3 analog outputs

iSeries: 16 digital inputs; 10 digital outputs; 6 analog outputs

Power Fail Relay Contact closure based signal relay to DAS, DCS, control room,

visible / audible alarm

Page 25: AQMS PRESENTATION THERMO

25

iSeries Measurement Electronics

All Electronics Board

• Upgraded electronics is necessary due to parts obsolescence/cost

• Snap-fit mounting connectors instead of screws to reduce assembly

time

Pressure and Flow Sensor Boards

• Upgraded electronics

• Removed pots/jumpers and replaced with software calibration

Permeation Oven

• Removed temperature control board, all temperature measurement

and control now handled by measurement interface board

• Kept same connectors to remain backwards compatible with C

Series.

Page 26: AQMS PRESENTATION THERMO

26

iSeries Measurement Electronics

Measurement Interface Boards

• Provides all power and connections to measurement system

• Different for each measurement technology

• 3 Connections to common electronics: 24VDC Power, 110VAC Power, RS-485

• Upgraded electronics, removed pots/jumpers where possible

• On-board processor handles data acquisition and low-level signal processing to/from:

• Up to 2 frequency inputs for primary measurements

• Up to 16 analog inputs for other measurements

• Up to 4 analog outputs for measurement system control

• Up to 1 SPI port for measurement system control

• Up to 2 PWM outputs & 1 bang/bang output for heater control

• Up to 16 solenoid outputs

• Up to 8 digital control outputs

• Up to 8 digital status inputs

Page 27: AQMS PRESENTATION THERMO

27

iSeries New Features

FEATURES BENEFITS

Enhanced Electronics Design Improved Serviceability

33% Reduction in Common PCBA

Analog Inputs External Sensor Signal Capture

Communication Protocol

Flexibility

TEI For Windows Support

Web Based (HTTP) Support

ModBus Support

Power Fail Relay Remote Power Failure Indication

Page 28: AQMS PRESENTATION THERMO

28

iSeries Platform – Other Features

Replaced side clips with captive thumb screws

Standard EIA rack mount as well as retrofit C-Series rack mount

Integral mounting tabs to reduce screws and assembly time

Mounted sub-assemblies on plates using alignment pins and

captive hardware to reduce assembly time and facilitate

outsourcing

Hinged panel for mounting measurement interface board which

allows greater access to components during assembly

Optimized component layout to facilitate one-step removal

process (reducing the need to remove other components).

Page 29: AQMS PRESENTATION THERMO

29

iSeries Common Electronics – I/O

2 RS-232/485 Serial Connectors

• Single port that may be daisy chained

• Software programmable for RS-232 or RS-485

• Software selectable baud rate

Ethernet Port

6 Analog Outputs

• Software selectable measurement channel and range

• User selectable 0-100mV,1V,5V,10V

16 Digital Inputs

• Software selectable triggers for measurement mode, calibrations, etc.

Power Fail Relay

• Both NC and NO contacts for connection versatility

Page 30: AQMS PRESENTATION THERMO

30

iSeries I/O

Digital Output Card (Standard)

• 10 Relays • Software selectable logic state (open or closed)

• Software selectable output data

Any alarm condition

Autorange status

Service mode

Units selection (ppb or ug/m3)

Zero/span/sample mode

I/O Expansion Card (Optional)

• 6 0-20/4-20mA Analog Outputs • Software selectable measurement channel and range

• Independent of voltage outputs

• 8 0-10V Analog Inputs • Input from external sensors

• Conversion table to read units other than volts

Page 31: AQMS PRESENTATION THERMO

31

iSeries Datalogging

I.B Data Logging

• LRECs and SRECs kept for backward compatibility.

• Alarm indicators kept for backward compatibility.

• User may select memory allocated for LRECs and SRECs in 10%

increments (no more 2K blocks).

• User may select what data is logged (up to 32 items each).

• 3MB FLASH allocated for logged data

• 42i default stores 25,000 LRECs (1 minute logging is 18 days of data). 42C stored

1,500 LRECs (1 minute logging is 25.5 hours of data).

• 42i default stores 96,000 SRECs (1 minute logging is 53 days of data). 42C stored

2,500 SRECs (1 minute logging is 42.6 hours of data).

• Worst-case 1 week of data with full LRECs & SRECs on 1 min logging.

• Best-case 100% LRECs date/time/flags/conc 190,000 records.

• Logged data may be viewed on front-panel by date or record num.

Page 32: AQMS PRESENTATION THERMO

32

i Port Software

Following settings of screen refresh values makes i port software

run better.

Suggested refresh values depending on the baud rate:

This can be changed in the preferences menu:

9600 = 30 sec

19200 = 15 sec

38400 = 10 sec

57600 = 8 sec

115200 = 5 sec

Page 33: AQMS PRESENTATION THERMO

33

iSeries I/O

Serial Port

• Two connectors to support daisy chaining (like C Series)

• DB9 Male Connectors, DTE configuration (like C Series)

• Software selectable for RS-232 or RS-485

• RS-232 pinouts identical to C Series

• RS-485 pinouts different from C Series!!!

• Software selectable baud rate (1200 – 115200 Baud)

• Software selectable communications protocol

• Thermo CLINK

• Thermo Streaming

• MODBUS RTU

Page 34: AQMS PRESENTATION THERMO

34

Updated Data Logging

Page 35: AQMS PRESENTATION THERMO

35

Allow Over/Under Range Outputs

Page 36: AQMS PRESENTATION THERMO

The world leader in serving science

SO2 Analysers

Model 43i

Page 37: AQMS PRESENTATION THERMO

37

“ Fluorescence “

- emission of light or Luminance from electronically excited state of a molecule

Sulphur dioxide absorbs light in three primary UV regions.

Out of these, the band between 190 nm and 230 nm is best suited for

fluorescence measurements.

Absorption of ultraviolet radiation by SO2 molecule yields an excited

Molecule SO2*. SO2* then releases its excess energy by fluorescence

at a different wavelength, typically between 240 - 420 nm

Measurement Principle

Page 38: AQMS PRESENTATION THERMO

38

Measurement Principle

Ia

Kf

SO2 + Hn1 SO2*

SO2* SO2 + Hn2

Ia = Io[I-e-{ax(SO2)}]

F = K(SO2)

Io = Incident Intensity

Ia = Intensity absorbed by SO2 Molecules

a = Absorption Coefficient of SO2

x = Optical path length

(SO2) = Concentration

Page 39: AQMS PRESENTATION THERMO

39

Measurement Principle

Page 40: AQMS PRESENTATION THERMO

40

The sample gas flows through a "Kicker" which removes hydrocarbons from

the gas stream through a selective permeation membrane.

The excitation light is produced by a pulsed xenon flash lamp and filtered by

a set of four interference mirrors (excitation filter) which transmit light

between 195 - 235 nm.

The emission filter is a combination of two band pass filters.

It selects light between 340 and 410 nm before the emitted radiation

reaches an orthogonally oriented photomultiplier tube.

A photo detector monitors the light from the lamp. Its signal is

compared to a reference signal and used to control the lamp intensity.

Theory of Operation

Page 41: AQMS PRESENTATION THERMO

41

Signal Processing Power Supply

Pressure

Transducer

Flow

Transducer

Sample

Span

Zero

Sample Pump

Exhaust

Capillary

PMT Detector Band-pass

Filter

U.V. Lamp

Span/Zero

Valve

Option

Photo-detector (feedback)

HC Kicker

Flow Schematics

Page 42: AQMS PRESENTATION THERMO

42

PHOTO MULTIPLIER TUBE

U.V. LIGHT

SOURCE

REFLECTIVE BANDPASS FILTERS

CONDENSING LENS

RELAY LENS

CIRCULAR BAFFLE

REACTION VOLUME

BANDPASS FILTER

PHOTO-DETECTO

R

ELECTRONICS

Pulsed source lamp

Reflective filtering

Circular baffle

Constant U.V. excitation level

MODEL 43i

Page 43: AQMS PRESENTATION THERMO

43

FEATURES BENEFITS

Pulsed Fluorescence

Design

No Consumables

Long Lamp Life

High Optical Density

Long Term Stability

Preferred Method

Worldwide

Hydrocarbon Kicker SO2 Specific Analysis

Selective Permeation

Removal

Flow Rate Independent

Design Enhancements

Page 44: AQMS PRESENTATION THERMO

44

FEATURES BENEFITS

Reflective U.V. Filtering

Wavelength Selectivity

Increased Sensitivity

Increased Stability

Circular Baffle Minimizes Stray Light

Interference

Thermally Controlled

Optical Bench

Drift Minimization

Improved

Reproducibility

Design Enhancements

Page 45: AQMS PRESENTATION THERMO

45

FEATURES BENEFITS

U.V. Reference Detector Continuous Monitoring

of Incident Light

Automatic Feedback

of Lamp Excitation

Energy

Design Enhancements

Page 46: AQMS PRESENTATION THERMO

46

Display Voltages of Main Board, Lamp & PMT Power

supplies

Temperature Display

Pressure Display

Flow Display

Lamp Intensity

Optical Span Test

Test Analog Outputs

Instrument Configuration

Digital Inputs

Relay States

Diagnostic Features

Page 47: AQMS PRESENTATION THERMO

47

Trouble Shooting Guide

Page 48: AQMS PRESENTATION THERMO

48

Trouble Shooting Guide

Page 49: AQMS PRESENTATION THERMO

49

Preventive Maintenance Calendar – 43i

Activity Frequency

Visual Inspection and Cleaning Once a Week

Sample Particulate Filter Replacement Once in 15 days

Fan Filter Inspection and Cleaning Once in 15 days

Signals Check Once in a Month

Calibration Check as per regulations

Capillary Inspection and Replacement Once in Six Months

Leak test Once in Six Months

Pump Rebuilding Once in Six Months

Check Flash Pump Once in a Year

Check or Replace Hydrocarbon Kicker Once in a Year

Page 50: AQMS PRESENTATION THERMO

50

Page 51: AQMS PRESENTATION THERMO

51

Model 450i Principle of Operation

Page 52: AQMS PRESENTATION THERMO

52

Page 53: AQMS PRESENTATION THERMO

53

Specifications

Page 54: AQMS PRESENTATION THERMO

54

Measurement Display

Page 55: AQMS PRESENTATION THERMO

55

Measurement Mode

Page 56: AQMS PRESENTATION THERMO

The world leader in serving science

Model 42i NO-NO2-NOX

NOX Analyser

Model 42i

Page 57: AQMS PRESENTATION THERMO

57

MODEL 42i Principle of Operation

Page 58: AQMS PRESENTATION THERMO

58

Model 42i Flow Schematic

Page 59: AQMS PRESENTATION THERMO

59

Measurement Calculations

Page 60: AQMS PRESENTATION THERMO

60

Specifications

Page 61: AQMS PRESENTATION THERMO

61

Specifications

Page 62: AQMS PRESENTATION THERMO

62

Single/Dual Range

Page 63: AQMS PRESENTATION THERMO

63

Auto Range

Page 64: AQMS PRESENTATION THERMO

64

Auto Range

Page 65: AQMS PRESENTATION THERMO

65

Gas Units

Page 66: AQMS PRESENTATION THERMO

66

Custom Ranges

Page 67: AQMS PRESENTATION THERMO

67

Averaging Time

Page 68: AQMS PRESENTATION THERMO

68

Auto/Manual Mode

Page 69: AQMS PRESENTATION THERMO

69

Ozonator Safety

Page 70: AQMS PRESENTATION THERMO

70

Preventive Maintenance

Page 71: AQMS PRESENTATION THERMO

71

Page 72: AQMS PRESENTATION THERMO

72

Page 73: AQMS PRESENTATION THERMO

73

Flow Schematics

Page 74: AQMS PRESENTATION THERMO

74

Theory of Operation

Page 75: AQMS PRESENTATION THERMO

75

Theory of Operation

Page 76: AQMS PRESENTATION THERMO

76

Specifications

Page 77: AQMS PRESENTATION THERMO

77

Page 78: AQMS PRESENTATION THERMO

78

Measurement Modes

NO Manual

NOX Manual

NT Manual

AUTO

NO

NOX

NT

NO

NO2

NH3

Page 79: AQMS PRESENTATION THERMO

79

Model 48i CO Analyser

Model 48i CO Analyzer

Page 80: AQMS PRESENTATION THERMO

80 15

iSeries Measurement Electronics –Gas Filter Correlation Analyzers (48i, 410i)

• Measurement Interface Board

– Combines motherboard, power supply board, bias supply, bench temperature control, and input board from C Series

• Preamp Board

– Software gain selection instead of resistor selection to simplify assembly

Page 81: AQMS PRESENTATION THERMO

81

Principle of Operation

Page 82: AQMS PRESENTATION THERMO

82

Model 48i Flow Schematic

Page 83: AQMS PRESENTATION THERMO

83

Preventive Maintenance

Preventive Maintenance Calender Model 48C

Sr Description Frequency Criteria

1 Visual Inspection and Cleaning Weekly Check dust particles or bad contacts

2 Clean or Replace Dust filters Weekly check particle deposits

3 Clean or Replace fan protection filters Weekly High internal temp of analyser

4 Clean Capillaries Once in a month dropped flow rates even with good pump

5 Clean Optics

Best results are obtained when

optics is cleaned prior to

calibration, typically once in six

months.

Cleanliness of the mirrors should be

checked anytime AGC intensity drops

below 200,000 Hz.

6 Replace IR Source

IR Source have a finite life, since it is

relatively inexpensive and easily

replacable, we recommend

replacing it annually.

There is no light output or even after

cleaning the optics IR intensity

remains lower than 100,000hz.is

7 System Leak Checks, replacement of Solenoid

valves Annually

even when pump diaphragm is good and

capillaries not blocked; if plugging

inlets takes more than 60 sec for

pressure value to drop below 250mm

Hg.

8 Pump rebuilding Annually even after cleaning Capillaries if flow is

below 0.350LPM

Caution: Optical mirrors are easy to scratch, this may lead to harm than any good.

Page 84: AQMS PRESENTATION THERMO

84

Troubleshooting

Here are a few of the best things to do/know when troubleshooting our I/R Instruments.

1. When you request AGC Intensity, the most critical Intensity to have is the AGC OFF Intensity.

This is obtained in the service menu under intensity check. It will first show you the AGC Intensity

then you can

toggle the AGC off. AGC off Intensity should be no lower than 150,000 Hz. 200,000 Hz is Typical

with the AGC On.

This must be obtained while flowing zero air.

2. When ever you get the background limit error, the first thing you should check is in service

mode go to Initial S/R Ratio.

If the unit is a 48C, the S/R Ratio must be between 1.13 - 1.19 with Zero air flowing.

The Initial S/R Ratio must be set to what the Current S/R Ratio is reading.

For Example: If the Current is 1.14 and the Initial is 1.17.

Change the initial S/R to be equal to the Current which is 1.14.

They should both now be reading 1.14 as an example.

When these two numbers are not close together, you will get the background error.

Note: If the Current S/R Ratio is out of our tolerance of 1.13 - 1.19, here are some things that

could make it low.

A: Dirty Gas Correlation Wheel

B: Wheel needs to be replaced.

C: The unit may have a purge housing around the wheel and the purge air is not CO Free.

D: A bad Interference Filter.

Page 85: AQMS PRESENTATION THERMO

85

Specification

Page 86: AQMS PRESENTATION THERMO

86

410i

Page 87: AQMS PRESENTATION THERMO

87

410i Schematics

Page 88: AQMS PRESENTATION THERMO

88

410i Specifications

Page 89: AQMS PRESENTATION THERMO

89

Online PM 10 / PM 2.5 / PM 1 / TSP Monitor b Attenuation

Series FH 62 C14

Page 90: AQMS PRESENTATION THERMO

90

USEPA Designated for PM10

CARB Designated for PM10 & PM2.5

Series FH 62 C14 Features and Benefits

• Automatic and adjustable filter changing for optimizing equilibrium and filter loading changes

• Accounts for Natural Radioactivity

• 1-yr data storage and minimum 6 month filter tape

• Insensitive to vibration and diurnal temperatures

• Mass measurement independent of aerosol properties

• Data Output via Analog, Serial, and Display

• Long term networking ability

• Microprocessor controlled, Quality Assurance Features

Page 91: AQMS PRESENTATION THERMO

91

Adjustable Inlet Heater (smart heater pending)

Series FH 62 C14 Features and Benefits

• Analog I/O Expansion Board: Can be used with meteorological sensors

• Filter Strip Printer for quasi-Chain of Custody

• Significant mass loading for post collection analysis

Page 92: AQMS PRESENTATION THERMO

92

TEOM® Series 1400a Ambient Particulate Monitor

Operating Principle Overview

Setting the standard for

particulate monitoring worldwide

Direct, continuous, filter-based

USEPA Equivalent Method for

PM-10: EQPM-1090-079

German TÜV Report for EN12341

European PM-10 standard

Page 93: AQMS PRESENTATION THERMO

93

TEOM Mass Transducer

Tapered element oscillates at its

natural frequency

Particulate matter collects on filter

continuously

Frequency decreases with

accumulation of mass

Direct relationship between mass and

frequency change

Tapered Element

Page 94: AQMS PRESENTATION THERMO

94

TEOM Monitor Mass Measurement Principle

M g Kf f

o

o

( )

1 1

1

2 2

Data averages and update intervals:

• 5-Minute Total Mass Average (every 2 sec)

• 10-Minute Rolling MC Average (every 2 sec)

• 30-Minute MC Average (every 30 minutes)

• 1-Hour MC Average (every 1 hour)

• 8-Hour MC Average (every 1 hour)

• 24-Hour MC Average (every 1 hour)

Page 95: AQMS PRESENTATION THERMO

95

1400A Mass Measurement

Instrument Performance (3 l/min, 1s,stable conditions)

Measurement Range: 0 to 5,000,000 µg/m³ (5 g/m³)

Resolution: 0.1 µg/m³

Precision: ±1.5 µg/m³ (1-hour ave), ±0.5 µg/m³ (24-hour ave)

Minimum Detectable Limit for Mass Measurement: 10

nanograms, 0.06 µg/m³ (1-hour ave)

Accuracy for Mass Measurement: ±0.75%

Page 96: AQMS PRESENTATION THERMO

96

1400A Specifications

Power Requirements

• Sensor and Control Units: 120 VAC/60 Hz: 1 A; 240 VAC/50

Hz: 0.5A

• Pump: 120 VAC/60 Hz: 4.25 A; 240 VAC/50 Hz: 2.25 A

Physical Dimensions:

• Base of TEOM Sensor Unit: W: 14" (36 cm) x D: 11" (28

cm) x H: 13“ (33) cm;

• Heated Air Inlet of TEOM Sensor Unit: H: 26" (66 cm) x

Diameter:3.5" (9 cm); Weight: 40 lb (18 kg)

• TEOM Control Unit: W: 17" (43 cm) x D: 18" (46 cm) x H: 9"

(22 cm) (rack mountable); Weight: 32 lb (15 kg)

Page 97: AQMS PRESENTATION THERMO

97

1400A Specifications

Operating Range:

• The temperature of the sampled air may vary between -

40 and 60C.

• The TEOM Sensor and Control Units must be weather

protected within the range of 2 to 40 °C. An optional

Complete Outdoor Enclosure provides complete

weather protection.

• Main Flow Rate: 0.5 to 4.0 l/min

• Auxiliary Flow Rate: 2.0 to 18.0 l/min

• Temperature of Mass Transducer: ambient to 70 °C

• Temperature of Internal Sample Tube: ambient to 70 °C

Page 98: AQMS PRESENTATION THERMO

98

8500 FDMS Typical Installation

Page 99: AQMS PRESENTATION THERMO

99

Mass Concentration Calculation

Based upon mass concentration (MC) measurements obtained during

the base and reference periods, the FDMS system updates a one-hour

average of the following results every six minutes :

Base mass concentration (Base MC) = PM concentration of the

particleladen sample stream (comparable to the Sample Equilibration

System at 30° C).

Reference mass concentration (Ref MC) = PM concentration of the

particle-free sample stream, after passing through the purge filter.

Mass concentration (MC) = Base mass concentration (Base MC)

adjusted by the reference mass concentration (Ref MC) — Base MC

(usually positive) minus Ref MC (negative when mass volatilizes from

the filter).

MC = Base MC - Ref MC

Page 100: AQMS PRESENTATION THERMO

100

Series 1405 DF

The Series 1405-DF Monitor is composed of two major

components (Figure 1-1):

the sample inlet assembly with virtual impactor and the

Series 1405-DF unit with FDMS system. The user

enters the system parameters into the Series 1405-DF

unit using a color touch screen that is located on the

front of the unit. Additionally, the system is furnished

with software for personal computers (PCs) to allow

the user to download data and update instrument

firmware. The instrument does not require a dedicated

computer to function in the field.

Page 101: AQMS PRESENTATION THERMO

101

TEOM Series 1405

The Series 1405-D Monitor is composed of two major components: the sample inlet assembly with virtual impactor and the Series 1405-D unit. The user enters the system parameters into the Series 1405-D unit using a color touch screen that is located on the front of the unit. Additionally, the system is furnished with software for personal computers (PCs) to allow the user to download data and update instrument firmware. The instrument does not require a dedicated computer to function in the field.

The sensor unit contains the two mass measurement hardware systems that continuously monitor both fine and coarse particulate. PM Coarse and PM fine particulate, split by a virtual impactor, each accumulate on the system’s exchangeable TEOM filters. By maintaining a flow rate of 1.67 l/ min through the coarse particulate flow channel and 3 l/min through the fine particulate channel, and measuring the total mass accumulated on each of the TEOM filters, the device can calculate the mass concentration of both the fine and coarse sample streams in near real-time.

The sampling system consists of a size-selective inlet, flow splitter

Page 102: AQMS PRESENTATION THERMO

102

Achievements in Air Quality monitoring

Experience of many years in supplying, maintenance of ambient

as well as stack monitoring stations.

ThermoFisher and Chemtrols have been awarded with the

biggest order for 61 nos. ambient monitoring stations in India by

NTPC.

Following monitors will be integrated in one station

SO2 Analyser : Model 43i

NOX Analyser : Model 42i

PM 10 Monitor : Model FH 62 C 14

CO2 Analyser : Model 410i

Multi gas Calibrator : Model 146i

With calibration gas cylinders, S/W and other accessories

Page 103: AQMS PRESENTATION THERMO

103

Bill Of Material for One AAQM Station

Analysers (NOX, SO2, CO, PM 10)

Multi Gas Calibrator (MGC)

Zero Air Generator required for MGC

Sample Handling System (Moisture Trap for Gas Analysers)

Sampling head for PM 10 Sampling

Calibration Gas Cylinders with Regulators

Data logging System (PC, Communication S/W, Telephone line

& Modem for Central Station Communication)

Rack, sample tubes, communication cables

Page 104: AQMS PRESENTATION THERMO

104

Continuous Ambient Air Quality

Monitoring Systems

Continuous Emission Monitoring Systems

Opacity/Dust

Hydrocarbon Analysers

Calibrators and Zero Air Generators

Pollution Monitoring Mobile Vans

Analyser Shelters

Turnkey Projects

Data Acquisition Systems for Ambient

& Stack Monitoring

Environment Instrumentation

Page 105: AQMS PRESENTATION THERMO

105

More than 25 dedicated Customer Support Personnel dedicated for

Environment Instrumentation Division

• Broad Functions of EI includes:

• Support from Conception to completion

• Assist Customer to Choose Right Solution

• Integration and Project Supply

• Mobile Van/ Fixed Shelter Design and Fabrication

• Quality Maintenance and Checks

• Approval by Customer / Third Party Inspection

• Prerequisites Readiness

• Installation and Commissioning.

• Customer Training

• AMC, CMC, RC/ ADHOC

Environment Instrumentation Division

Page 106: AQMS PRESENTATION THERMO

106

AQM/CEM

Central Data

Station

ST 6

EnviDAS Software

Plant DCS/PLC

And Display Board

COMMUNICATION

RS232/RS 485 /GSM or

Tele phone Modem

ST 4 ST 1 ST 3

ST 2

ST 5

Page 107: AQMS PRESENTATION THERMO

107

Thank You!

Naveen kumar Engineer-EID

Delhi - NOIDA

Cell : 9690769283

www.thermo.com/air